FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0

MDR report key: 7711073 · Received July 23, 2018

Report

Report Number
3007042319-2018-03185
Event Type
Malfunction
Date Received
July 23, 2018
Date of Event
June 29, 2018
Report Date
December 27, 2018
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
PMA / PMN Number
P100047
Removal / Correction Number
Z-1903-2018
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: ONE CONTROLLER (CON302502) AND SIX BATTERIES (BAT214411, BAT214261, BAT214101, BAT214115, BAT214412, BAT214338) WERE RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED CONTROLLER AND BATTERIES REVEALED THAT THE DEVICES PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER CONTAINED A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15-MINUTE INTERVAL. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS INVOLVING BAT214261, BAT214101, BAT214115, BAT214412, BAT214411, AND BAT214338, AS WELL AS A PREMATURE POWER SWITCHING EVENT THAT WAS DUE TO A COMMUNICATION ERROR INVOLVING BAT214411. LOG FILE ANALYSIS ALSO REVEALED THREE CONTROLLER POWER UP EVENTS, WITH ASSOCIATED MOTOR START EVENTS, ON 30MAY2018 AT 20:51:52, 27JUNE2018 AT 13:47:04, AND 29JUNE2018 AT 11:58:18, INDICATING LOSSES OF POWER TO THE CONTROLLER. SEVERAL MOMENTARY DISCONNECTIONS WERE RECORDED LEADING UP TO THE FIRST AND SECOND LOSSES OF POWER. THE CONTROLLER WAS WITHOUT POWER FOR 11 SECONDS, 9 SECONDS, AND 12 SECONDS, RESPECTIVELY. AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE REPORTED POWER SWITCHING EVENT CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS AND COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. A POSSIBLE ROOT CAUSE OF THE LOSSES OF POWER CAN BE ATTRIBUTED TO A DISCONNECTION OF BOTH POWER SOURCES AND/OR TO AN INTERMITTENT DISCONNECTION ON ONE OR BOTH POWER SOURCES. INTERNAL INVESTIGATIONS WERE INITIATED TO CAPTURE EVENTS INVOLVING THE CONTROLLER LOSING POWER AND TO EVALUATE MOMENTARY DISCONNECTIONS. ADDITIONAL PRODUCTS: BATTERY BAT214411 H3: YES H6 FDA METHOD CODE(S): 10, 4112 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 12 BATTERY BAT214261 H3: YES H6 FDA METHOD CODE(S): 10, 4112 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 12, 4307 BATTERY BAT214101 H3: YES H6 FDA METHOD CODE(S): 10, 4112 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 12 BATTERY BAT214115 H3: YES H6 FDA METHOD CODE(S): 10, 4112 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 12 BATTERY BAT214412 H3: YES H6 FDA METHOD CODE(S): 10, 4112 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 12 BATTERY BAT214338 H3: YES H6 FDA METHOD CODE(S): 10, 4112 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 12 MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. OTHER DEVICES INVOLVED IN THIS EVENT: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2017-02-26 RETURN DATE: 2018-07-09 DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN MFG DATE: 2016-02-28 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2017-02-26 RETURN DATE: 2018-07-09 DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN MFG DATE: 2016-02-28 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2017-02-26 RETURN DATE: 2018-07-09 DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN MFG DATE: 2016-02-28 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2017-02-26 RETURN DATE: 2018-07-09 DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN MFG DATE: 2016-02-28 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2017-02-26 RETURN DATE: 2018-07-09 DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN MFG DATE: 2016-02-28 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2017-02-26 RETURN DATE: 2018-07-09 DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN MFG DATE: 2016-02-28 (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CONTROLLER AND BATTERIES EXHIBITED POWER SWITCHING WITH A SUBSEQUENT PUMP STOP. THE CONTROLLER AND BATTERIES WERE EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
551612 HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0 VENTRICULAR (ASSISST) BYPASS DSQ HEARTWARE, INC. 1420

Patients

Seq Age Sex Outcome Treatment
1 67 YR