FDA Adverse Event
Malfunction
Summary report: N
APEX K2 BROACH
MDR report key: 770919
·
Received August 7, 2006
Report
- Report Number
- 1226188-2006-00004
- Event Type
- Malfunction
- Date Received
- August 7, 2006
- Date of Event
- May 5, 2006
- Report Date
- June 27, 2006
- Manufacturer
- OMNI LIFE SCIENCE, INC.
- Product Code
- HTQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON EVAL OF THE RETURNED DEVICE, THE TIP HAD BROKEN OFF AT THE FIRST LARGE TOOTH. WE WILL BE MOVING TO A FINER PITCH TOOTH FOR FUTURE ORDERS.
Description of Event or Problem · 1
THE TIP OF A K2 BROACH, BROKE OFF DURING TOTAL HIP REPLACEMENT SURGERY. THE BROKEN PART WAS SUCCESSFULLY RETRIEVED BY THE SURGEON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | APEX K2 BROACH | SURGICAL INSTRUMENT FOR THA | HTQ | OMNI LIFE SCIENCE, INC. | SIZE 2 BROACH | L0405 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Required Intervention |