FDA Adverse Event Malfunction Summary report: N

APEX K2 BROACH

MDR report key: 770919 · Received August 7, 2006

Report

Report Number
1226188-2006-00004
Event Type
Malfunction
Date Received
August 7, 2006
Date of Event
May 5, 2006
Report Date
June 27, 2006
Manufacturer
OMNI LIFE SCIENCE, INC.
Product Code
HTQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON EVAL OF THE RETURNED DEVICE, THE TIP HAD BROKEN OFF AT THE FIRST LARGE TOOTH. WE WILL BE MOVING TO A FINER PITCH TOOTH FOR FUTURE ORDERS.

Description of Event or Problem · 1

THE TIP OF A K2 BROACH, BROKE OFF DURING TOTAL HIP REPLACEMENT SURGERY. THE BROKEN PART WAS SUCCESSFULLY RETRIEVED BY THE SURGEON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 APEX K2 BROACH SURGICAL INSTRUMENT FOR THA HTQ OMNI LIFE SCIENCE, INC. SIZE 2 BROACH L0405

Patients

Seq Age Sex Outcome Treatment
1 NA Required Intervention