OPTIFLUX 180NR DIALYZER FINISHED ASSMBLY
Report
- Report Number
- 1713747-2018-00259
- Event Type
- Malfunction
- Date Received
- July 21, 2018
- Date of Event
- July 13, 2018
- Report Date
- September 12, 2018
- Manufacturer
- OGDEN MANUFACTURING PLANT
- Product Code
- KDI
- UDI-DI
- 00840861100125
- PMA / PMN Number
- K002761
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
PMA/510(K) TO K002761.
PLANT INVESTIGATION: THE REPORTED COMPLAINT WAS NOT CONFIRMED AS THE COMPLAINT DEVICE WAS NOT RETURNED FOR MANUFACTURER EVALUATION. A PRODUCTION RECORDS REVIEW WAS PERFORMED ON THE REPORTED LOT. AN INVESTIGATION OF THE DEVICE HISTORY RECORDS (DHR) WAS CONDUCTED BY THE MANUFACTURER. THERE WAS ONE APPROVED TEMPORARY DEVIATION NOTICE (DN) REPORTED ON THE LOT WHICH WAS UNRELATED TO THE COMPLAINT EVENT. THERE WAS NO INDICATION OF PRODUCT NONACCEPTANCE, DEVIATION, NON-CONFORMANCE, REWORK, LABELING OR PROCESS CONTROL FAILURE DURING THE MANUFACTURING PROCESS WHICH COULD BE ASSOCIATED WITH THE REPORTED EVENT. THE LOT MET ALL RELEASE CRITERIA. A DEFINITIVE CONCLUSION REGARDING THE COMPLAINT INCIDENT CANNOT BE REACHED WITHOUT PHYSICAL EXAMINATION OF THE ACTUAL DEVICE. THEREFORE, THE COMPLAINT IS NOT CONFIRMED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
CORRECTED CATALOG NUMBER TO 0500318E, LOT NUMBER TO 18CU0515, AND UDI TO (B)(4).
THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.
A USER FACILITY CLINICAL MANAGER REPORTED THAT A DIALYZER BLOOD LEAK OCCURRED IMMEDIATELY AFTER THE INITIATION OF THE PATIENT¿S HEMODIALYSIS (HD) TREATMENT. THE BLOOD LEAK WAS NOTED AS BEING AN INTERNAL BLOOD LEAK, AND WAS VISUALLY OBSERVED. THE MACHINE, A FRESENIUS 2008T MACHINE, ALARMED APPROPRIATELY WITH A BLOOD LEAK ALARM. THE PATIENT¿S ESTIMATED BLOOD LOSS (EBL) WAS APPROXIMATELY 300ML. THERE WAS NO PATIENT INJURY, ADVERSE EVENTS, OR MEDICAL INTERVENTION REQUIRED AS A RESULT OF THIS EVENT. THE PATIENT WAS RESTARTED ON A NEW MACHINE AND TREATMENT COMPLETED SUCCESSFULLY WITH NEW SUPPLIES. THE COMPLAINT DEVICE WAS REPORTED TO BE AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 550510 | OPTIFLUX 180NR DIALYZER FINISHED ASSMBLY | DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM | KDI | OGDEN MANUFACTURING PLANT | 18CU0515 | 00840861100125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |