FDA Adverse Event Malfunction Summary report: N

MAVERICK2 MONORAIL PTCA CATHETER

MDR report key: 770894 · Received August 16, 2006

Report

Report Number
6000093-2006-01534
Event Type
Malfunction
Date Received
August 16, 2006
Date of Event
June 29, 2006
Report Date
July 20, 2006
Manufacturer
BOSTON SCIENTIFIC
Product Code
LOX
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RETURNED PRODUCT ANALYSIS COULD NOT VERIFY THE DIFFICULTY STATE IN THE COMPLAINT. THE BALLOON CATHETER WITH THE BALLOON IN A DEFLATED STATE WAS RECEIVED IN GOOD CONDITION WITH NO DAMAGE OBSERVED. VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE DID NOT REVEAL ANY DEFECTS OR DAMAGE TO THE DEVICE THAT WOULD HAVE CONTRIBUTED TO THE BALLOON LEAK DIFFICULTIES. THE CATHETER WAS PURGED COMPLETELY OF AIR AND THE BALLOON WAS INFLATED WITH WATER TO ITS RATED BURST PRESSURE. THE CATHETER WAS PRESSURIZED WITHOUT SIGNIFICANT LOSS OF PRESSURE. WHEN THE CATHETER WAS SUBJECTED TO A VACUUM, THE BALLOON DEFLATED NORMALLY AND THERE WAS NO COMPROMISE OF THE GENERATED VACUUM. THE CAUSE OF THE DIFFICULTIES ENCOUNTERED DURING THE PROCEDURE COULD NOT BE DETERMINED. THE MANUFACTURING RECORDS FOR TOP ASSEMBLY BATCH 7852136 HAVE BEEN REVIEWED AND NO ISSUES OR DISCREPANCIES WERE FOUND. THIS RECORDS REVIEW CONFIRMS THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PERFORMANCE SPECIFICATIONS. THERE ARE NO SIMILAR COMPLAINTS OF THIS NATURE RELATED TO THIS BATCH NUMBER. NO PROBLEM WAS FOUND WITH THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO A POBA PROCEDURE, THE MAVERICK2 MONORAIL BALLOON LEAKED AND A PINHOLE WAS NOTED DURING PREPARATION. THE LESION TO BE TREATED IS UNK. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT INJURIES OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAVERICK2 MONORAIL PTCA CATHETER PTCA CATHETER LOX BOSTON SCIENTIFIC 20MM X 3.0MM 7852136

Patients

Seq Age Sex Outcome Treatment
1 65 YR