MENTOR SILTEX ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2018-04365
- Event Type
- Injury
- Date Received
- July 19, 2018
- Date of Event
- March 9, 2018
- Report Date
- June 21, 2018
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001751
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
ON 8/15/2018, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ADDITIONAL INFORMATION INDICATED THAT THE RIGHT REPLACEMENT SERIAL NUMBER IS (B)(4), AND THE CORRECT REPLACEMENT LOT NUMBER IS 7438363. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
DEVICE EVALUATION SUMMARY: UPON RECEIPT BY MENTOR, THE DEVICE CONTAINED NO FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE. YELLOW MATERIAL WAS OBSERVED ON THE SHELL SURFACE. DURING VISUAL EXAMINATION OF THE DEVICE, PRODUCT EVALUATION TEAM OBSERVED TWO RENTS ON THE ANTERIOR VIEW MEASURING 1.8 AND 0.6 CM. MICROSCOPIC EXAMINATION OF THE RENT A EDGES REVEALED PARALLEL STRIATIONS WHICH ARE SIMILAR TO MARKINGS MADE BY A SHARP INSTRUMENT PERFORATING SILICONE MATERIAL. MICROSCOPIC EXAMINATION OF THE RENT B EDGES GAVE NO INDICATION AS TO ITS CAUSE. NO OTHER ANOMALIES WERE FOUND. MENTOR PERFORMS 100% INSPECTION AND TESTING OF ALL DEVICES PRIOR TO RELEASE; PRODUCT EVALUATION TEAM CONCLUDED THAT THE RENTS OCCURRED SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. THE COMPLAINT OF DEFLATION WAS CONFIRMED SINCE TWO RENTS WERE FOUND ON THE DEVICE. MICROSCOPIC EXAMINATION REVEALED PARALLEL STRIATIONS ON THE EDGES OF THE RENT A. THIS TYPE OF STRIATIONS IS MORE CONCLUSIVE TO SHARP INSTRUMENT DAMAGE RATHER THAN SHELL FAILURE DUE TO WEAR. A SECOND MICROSCOPIC EXAMINATION OF THE RENT B EDGES WAS PERFORMED, AND IT DID NOT PROVIDE CONCLUSIVE EVIDENCE OF WHAT COULD BE THE ROOT CAUSE. FURTHERMORE, AT THIS POINT IT IS ALSO NOT POSSIBLE TO DETERMINE THE ORIGIN OF THE YELLOW MATERIAL OBSERVED. MENTOR PERFORMS 100% INSPECTION AND TESTING OF ALL DEVICES PRIOR TO RELEASE; PRODUCT EVALUATION TEAM CONCLUDED THAT THE RENTS OCCURRED SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. MENTOR PERFORMS 100% INSPECTION AND TESTING OF ALL DEVICES PRIOR TO RELEASE; PRODUCT EVALUATION TEAM CONCLUDED THAT THE RENTS OCCURRED SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. ALL THE IMPLANTS ARE 100% INSPECTED BEFORE LEAVING THE FACILITY. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CONCOMITANT PRODUCTS: LEFT MENTOR SILTEX ROUND MODERATE PROFILE 225CC SALINE BREAST IMPLANT, CATALOG NUMBER 3542630, LOT NUMBER 5527907. MANUFACTURER¿S REFERENCE NUMBER: (B)(4). REPORTER TELEPHONE NUMBER: 02-2094-6588
IT WAS REPORTED THAT A (B)(6) FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION N WITH MENTOR SILTEX ROUND MODERATE PROFILE 225CC SALINE BREAST IMPLANTS THAT BILATERALLY DEFLATED AFTER IMPLANTATION. THE ISSUE WAS DIAGNOSED BY ULTRASONOGRAPHIC EXAMINATION, AS A RESULT, PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH CATALOG NUMBER 3502250, LOT NUMBER 7424431 ON (B)(6) 2018. THIS REPORT IS FOR THE RIGHT IMPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 545552 | MENTOR SILTEX ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 5557211 | 00081317001751 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Required Intervention |