ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2018-13072
- Event Type
- Malfunction
- Date Received
- July 18, 2018
- Report Date
- July 1, 2018
- Manufacturer
- ANIMAS LLC
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
OF THE 40 ALLEGATIONS OF A MALFUNCTION, 30 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 8 WERE FOUND TO BE MALFUNCTIONING DUE TO "REASON FOR FAILURE."
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01-OCT-2018 WITH THE FOLLOWING FINDINGS: IN QUARTER 3 OF 2018, THERE WERE 3 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.
THIS REPORT SUMMARIZES 40 MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGES FROM 5-74 YEARS OF AGE AND BETWEEN 20 AND 220 POUNDS. OF THE REPORTED PATIENTS, 18 WERE MALE AND 22 WERE FEMALE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 541453 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |