FDA Adverse Event
Other
Summary report: N
ZODIAC STAINLESS STEEL SPINAL FIXATION SYSTEM
MDR report key: 769736
·
Received September 22, 2006
Report
- Report Number
- 2027467-2006-00013
- Event Type
- Other
- Date Received
- September 22, 2006
- Report Date
- September 22, 2006
- Manufacturer
- ALPHATEC SPINE, INC.
- Product Code
- KWP
- PMA / PMN Number
- K051286
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ADDITIONAL PART INVOLVED: 8.5MM X 70MM LENGTH MONOAXIAL BONE SCREW PART NUMBER 82285-70. THE OFFSET CONNECTOR IS DISENGAGED FROM THE ENTIRE CONSTRUCT AT THE ILIAC BONE JUNCTION. THE DISENGAGED PARTS REMAIN IN THE PATIENT. REGULATORY AFFAIRS WAS MADE AWARE OF INCIDENT BY SURGEON WHO PERFORMED SURGERY.
Description of Event or Problem · 1
THE OFFSET CONNECTOR IS DISENGAGED FROM THE ENTIRE CONSTRUCT AT THE ILIAC BONE JUNCTION. THE DISENGAGED PARTS REMAIN IN THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ZODIAC STAINLESS STEEL SPINAL FIXATION SYSTEM | ZODIAC SS OFFSET CONNECTORS | KWP | ALPHATEC SPINE, INC. | 82006-20 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR |