FDA Adverse Event Injury Summary report: N

ANASTOCLIP CLIP APPLIER

MDR report key: 769457 · Received October 12, 2006

Report

Report Number
MW1040671
Event Type
Injury
Date Received
October 12, 2006
Date of Event
April 26, 2006
Report Date
September 20, 2006
Manufacturer
*
Product Code
GDO
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

BASILIC VEIN TRANSPOSTION WAS PERFORMED BY DR. HEMOSTASIS WAS ACHEIVED AT PRIMARY PLACEMENT USING PROLENE STAY STUTURE SNA LARGE VCA CLIPS. PT WAS ADMITTED TO ER AT ANOTHER HOSP DUE TO COMPLICATIONS OF THE CASE. THREE TO FOUR CLIPS ON THE FRONT WALL HAD SLIPPED OFF AND THE PT WAS BLEEDING OUT. THE PT WAS SEEN THAT NIGHT BY DR FOR AN EMERGENCY REVISION USING SUTURE FOR ANASTOMOSIS. DR PLACED ABOUT 15 CLIPS ON THE FRONT AND BACK WALL OF THE ANASTOMOSIS. DEVICE WAS NOT RETURNED FOR EVALUATION. LEMAITRE REPRESENTATIVE WENT TO THE HOSP AFTER THE INCIDENT AND SPOKE WITH THE PHYSICIAN. THE PHYSICIAN NOTED THAT HE MAY HAVE MISPLACED THE CLIPS RESULTING IN THE CLIPS COMING OFF. LEMAITRE REP SUGGESTED ATTENDING TRAINING SESSIONS THAT ARE OFFERED BY LEMAITRE VASCULAR, INC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANASTOCLIP CLIP APPLIER CLIP APPLIER GDO * NA ANC1027

Patients

Seq Age Sex Outcome Treatment
1 * Required Intervention