S100 OPMI PICO
Report
- Report Number
- 9615010-2018-00009
- Event Type
- Malfunction
- Date Received
- July 13, 2018
- Date of Event
- June 4, 2018
- Report Date
- June 13, 2018
- Manufacturer
- CARL ZEISS MEDITEC AG (CARL ZEISS MEDITEC AG (OBER
- Product Code
- HRM
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
DURING THE MANUFACTURER'S INVESTIGATION, IT WAS DISCOVERED THAT THE DEVICE HAD BEEN MODIFIED WITHOUT AUTHORIZATION FROM THE MANUFACTURER. THE S100 STAND (WHICH IS NORMALLY ATTACHED TO AN OPMI PICO HEAD) WAS FOUND ATTACHED TO THE HEAD FROM A DIFFERENT ZEISS MICROSCOPE - AN OPMI 1-FC. IN ADDITION, THE CASTORS ATTACHED TO THE BASE WERE FOUND TO BE NON-ZEISS SUPPLIED CASTORS. THE USER MANUAL FOR THE S100 CONTAINS CLEAR INSTRUCTIONS THAT ANY MODIFICATION OF THE SYSTEM IS FORBIDDEN. THE USER MANUAL ALSO STATES THAT ANY REPAIR WORK PERFORMED BY UNTRAINED AND/OR NON-CERTIFIED PERSON(S) IS PROHIBITED. ALSO, THE RECOMMENDED MAINTENANCE SCHEDULE FOR THIS DEVICE IS ANNUAL SERVICING; HOWEVER, NO SERVICE/MAINTENANCE RECORD FOR THIS DEVICE HAS BEEN FOUND BY THE MANUFACTURER. CARL ZEISS LTD. COMMUNICATED TO THE CUSTOMER TO TAKE THE MICROSCOPE OUT OF SERVICE.
A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT WHEN THE BRAKE WAS RELEASED ON THE S100 MICROSCOPE ARM, THE BALANCE FAILED AND THE MICROSCOPE HEAD FELL. IT WAS REPORTED BY THE CUSTOMER THAT THE PIVOT ON THE ARM'S PNEUMATIC SPRING HAD DISENGAGED FROM ITS MOUNT, AND HENCE WHEN THE BRAKE WAS ADJUSTED, THE MICROSCOPE HEAD FELL AND WAS CAUGHT BY THE USER. THE USER HAD TO CATCH THE HEAD TO AVOID HARM TO THE PATIENT. THE PNEUMATIC ARM IS HELD IN PLACE SOLELY BY TENSION AND IS NOT CAPTIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 527608 | S100 OPMI PICO | OPMI PICO | HRM | CARL ZEISS MEDITEC AG (CARL ZEISS MEDITEC AG (OBER | S100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |