FDA Adverse Event
Malfunction
Summary report: N
T:SLIM G5 SYSTEM
MDR report key: 7677815
·
Received July 11, 2018
Report
- Report Number
- 3013756811-2018-23261
- Event Type
- Malfunction
- Date Received
- July 11, 2018
- Date of Event
- June 18, 2018
- Report Date
- July 11, 2018
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OYC
- UDI-DI
- 00853052007257
- PMA / PMN Number
- P140015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE X2 PUMP USER GUIDE STATES DO NOT REUSE CARTRIDGES OR USE CARTRIDGES OTHER THAN THOSE MANUFACTURED BY TANDEM DIABETES CARE, INC. USE OF CARTRIDGES NOT MANUFACTURED BY TANDEM DIABETES CARE, INC. OR REUSE OF CARTRIDGES MAY RESULT IN OVER DELIVERY OR UNDER DELIVERY OF INSULIN. THIS CAN CAUSE VERY LOW OR VERY HIGH BLOOD GLUCOSE. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED AFTER FILLING A RE-USED CARTRIDGE WITH AROUND 105 UNITS OF INSULIN. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 308 MG/DL. REPORTEDLY, A NEW CARTRIDGE WAS LOADED SUCCESSFULLY AND INSULIN DELIVERY WAS RESUMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 520153 | T:SLIM G5 SYSTEM | CONTINUOUS GLUCOSE MONITOR | OYC | TANDEM DIABETES CARE | 1000096 | 00853052007257 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |