FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 7671414 · Received July 9, 2018

Report

Report Number
1645337-2018-04100
Event Type
Injury
Date Received
July 9, 2018
Date of Event
June 11, 2018
Report Date
June 13, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001294
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION SUMMARY: UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE OR ON THE SHELL SURFACE. PRODUCT EVALUATION TEAM DISCOVERED A RENT MEASURING APPROXIMATELY 5.2 CM WITHIN A CREASE ON THE ANTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE. THE DHR WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH MENTOR SMOOTH ROUND MODERATE PROFILE 475CC SALINE BREAST IMPLANTS WHICH THE RIGHT SIDE DEFLATED AFTER IMPLANTATION. THE ISSUE WAS CONFIRMED BY PHYSICAL EXAMINATION. AS A RESULT PATIENT HAD THE DEVICE REMOVED AND REPLACED WITH ANOTHER DEVICE CATALOG NUMBER 3501680, LOT NUMBER 6920283, SERIAL NUMBER (B)(4) ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
511765 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5978601 00081317001294

Patients

Seq Age Sex Outcome Treatment
1 36 YR Required Intervention