FDA Adverse Event Malfunction Summary report: N

SERVO-U

MDR report key: 7665854 · Received July 6, 2018

Report

Report Number
8010042-2018-00324
Event Type
Malfunction
Date Received
July 6, 2018
Date of Event
April 6, 2018
Report Date
September 13, 2018
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K151814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

THE VENTILATOR WAS INVESTIGATED ON SITE AND THE O2 GAS MODULE WAS REPLACED AND RETURNED FOR INVESTIGATION. THE O2 GAS MODULE REGULATES THE INSPIRATORY OXYGEN GAS FLOW TO THE PATIENT. THE REPORTED FAILURE IN THE SUBTEST FLOW TRANSDUCER TEST IN PRE-USE CHECK WAS NOT REPRODUCED DURING TEST OF THE RETURNED O2 GAS MODULE IN A VENTILATOR. THE OXYGEN GAS MODULE WAS FOUND TO DELIVER MORE OXYGEN GAS THAN SPECIFICATION DURING TEST IN THE PRODUCTION TEST SYSTEM FOR GAS MODULES. THE ELECTRONICS INSIDE THE O2 GAS MODULE CAUSED THE DEVIATION. IF THIS FAULT HAPPENS DURING VENTILATION, THE PATIENT MAY RECEIVE AN AMOUNT OF OXYGEN IN THE GAS MIXTURE THAT DEVIATE FROM THE EXPECTED. FLOW AND PRESSURE MAY ALSO DEVIATE FROM THE EXPECTED. ALARMS WILL BE ACTIVATED IF THE FAILURE OCCURS DURING VENTILATION. GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. (B)(4)

Description of Event or Problem · 0

MANUFACTURER REF. #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
510243 SERVO-U VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB

Patients

Seq Age Sex Outcome Treatment
1