BD VACUTAINER® CAT (CLOT ACTIVATOR TUBE) PLUS BLOOD COLLECTION TUBE
Report
- Report Number
- 9617032-2018-00295
- Event Type
- Malfunction
- Date Received
- July 3, 2018
- Date of Event
- June 13, 2018
- Report Date
- August 17, 2018
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER
Narratives
CORRECTION: DUE TO AN IT ISSUE BEGINNING ON 7/3/2018, PREVIOUSLY FILED EMDRS DID NOT CONTAIN REQUIRED FIELDS. THIS SUPPLEMENTAL EMDR IS FILED TO PROVIDE THE FOLLOWING OMITTED FIELDS: EVENT ATTRIBUTED TO: OTHER. DEVICE SINGLE USE?: NO. DEVICE RETURNED TO MANUFACTURER: NO. INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, EVALUATION AND TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND UNDERFILL WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. BASED ON EVALUATION OF THE CUSTOMER SAMPLES AND PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. A CAPA HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.
IT WAS REPORTED THAT WHEN USING A BD VACUTAINER® CAT (CLOT ACTIVATOR TUBE) PLUS BLOOD COLLECTION TUBE, THE SAMPLE UNDERFILLED. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION.
DATE OF EVENT: UNKNOWN. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8043505, MEDICAL DEVICE EXPIRATION DATE: 2020-02-29, DEVICE MANUFACTURE DATE: 2018-02-12, MEDICAL DEVICE LOT #: 7275913, MEDICAL DEVICE EXPIRATION DATE: 2019-09-30, DEVICE MANUFACTURE DATE: 2017-10-02. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHEN USING A BD VACUTAINER® CAT (CLOT ACTIVATOR TUBE) PLUS BLOOD COLLECTION TUBE, THE SAMPLE UNDERFILLED. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 500016 | BD VACUTAINER® CAT (CLOT ACTIVATOR TUBE) PLUS BLOOD COLLECTION TUBE | BLOOD COLLECTION TUBE | JKA | BECTON, DICKINSON AND COMPANY (BD) | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |