HIP END EFFECTOR, VARIABLE ANGLE
Report
- Report Number
- 3005985723-2018-00399
- Event Type
- Malfunction
- Date Received
- July 3, 2018
- Date of Event
- May 1, 2018
- Report Date
- July 12, 2018
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K121064
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 AND FINAL REPORT SUBMITTED TO UPDATE MANUFACTURER NAME, CITY AND STATE, MFR SITE, PMA/510K, IF FOLLOW-UP, WHAT TYPE AND DEVICE EVALUATED BY MFR BASED ON THE RESULTS OF INVESTIGATION. REPORTED EVENT: MICS WAS NOT WORKING, HAD TO REPLACE BEFORE REAMING, LATER IN THE CASE INLINE OFFSET IMPACTOR STUCK IN EE AFTER IMPACTION. THA PROCEDURE DELAYED A FEW MINUTES TO OPEN UP ANOTHER MICS; OTHERWISE, NO FURTHER DELAY FOR THE STUCK IMPACTOR. SURGICAL DELAY WAS ABOUT 5 MINUTES. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION SHOWS THAT THE 202857 DOWEL PIN IS MISSING FROM THE SIDE NEXT TO THE SERIAL NUMBER. SEE ATTACHMENTS FOR IMAGES OF THE INSTRUMENT. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED . FUNCTIONAL INSPECTION: FUNCTIONAL INSPECTION SHOWS THE DEVICE FUNCTIONS AS EXPECTED AND IS ABLE TO ACCEPT A KNOWN TO BE GOOD 209830 RIO SHELL IMPACT-OFFSET TRIDENT PST, BUT IT IS MISSING THE DOWEL PIN STATED ABOVE. MATERIAL ANALYSIS: MATERIAL ANALYSIS WAS COMPLETED AS PART OF THE ROOT CAUSE INVESTIGATION OF CAPA: 1450905. DEVICE HISTORY REVIEW: REVIEW OF THE PRODUCT HISTORY RECORDS INDICATE 20 DEVICES WERE MANUFACTURED UNDER LOT NO: 19060214 AND ACCEPTED INTO FINAL STOCK ON 06/26/2014 . NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N: 206967, LOT NUMBER: 19060214 SHOWS 05 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. THESE COMPLAINTS ARE: (B)(4). CONCLUSIONS: ALLEGED FAILURE CONFIRMED. SEE ATTACHED IMAGE. ROOT CAUSE CAN NOT BE DETERMINED. CORRECTIVE ACTION/PREVENTIVE ACTION: CAPA: 1450905 WAS INITIATED BASED ON OTHER COMPLAINTS REPORTING THIS ISSUE.
MICS WAS NOT WORKING, HAD TO REPLACE BEFORE REAMING, LATER IN THE CASE INLINE OFFSET IMPACTOR STUCK IN EE AFTER IMPACTION. THA PROCEDURE DELAYED A FEW MINUTES TO OPEN UP ANOTHER MICS; OTHERWISE, NO FURTHER DELAY FOR THE STUCK IMPACTOR. SURGICAL DELAY WAS ABOUT 5 MINUTES. PER THE INVESTIGATION, A PIN WAS MISSING FROM THE HIP EE.
REPORTED EVENT: MICS WAS NOT WORKING, HAD TO REPLACE BEFORE REAMING, LATER IN THE CASE INLINE OFFSET IMPACTOR STUCK IN EE AFTER IMPACTION. THA PROCEDURE DELAYED A FEW MINUTES TO OPEN UP ANOTHER MICS; OTHERWISE, NO FURTHER DELAY FOR THE STUCK IMPACTOR. SURGICAL DELAY WAS ABOUT 5 MINUTES. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION- VISUAL INSPECTION SHOWS THAT THE 202857 DOWEL PIN IS MISSING FROM THE SIDE NEXT TO THE SERIAL NUMBER. SEE ATTACHMENTS FOR IMAGES OF THE INSTRUMENT. DIMENSIONAL INSPECTION- DIMENSIONAL INSPECTION WAS NOT COMPLETED . FUNCTIONAL INSPECTION- FUNCTIONAL INSPECTION SHOWS THE DEVICE FUNCTIONS AS EXPECTED AND IS ABLE TO ACCEPT A KNOWN TO BE GOOD 209830 RIO SHELL IMPACT-OFFSET TRIDENT PST, BUT IT IS MISSING THE DOWEL PIN STATED ABOVE. MATERIAL ANALYSIS- MATERIAL ANALYSIS WAS COMPLETED AS PART OF THE ROOT CAUSE INVESTIGATION OF CAPA (B)(4). DEVICE HISTORY REVIEW: REVIEW OF THE PRODUCT HISTORY RECORDS INDICATE 20 DEVICES WERE MANUFACTURED UNDER LOT NO 19060214 AND ACCEPTED INTO FINAL STOCK ON 06/26/2014 . NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N (B)(4), LOT NUMBER 19060214 SHOWS 05 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. THESE COMPLAINTS ARE: (B)(4). CONCLUSIONS: ALLEGED FAILURE CONFIRMED. SEE ATTACHED IMAGE. ROOT CAUSE CAN NOT BE DETERMINED. CORRECTIVE ACTION/PREVENTIVE ACTION: CAPA (B)(4) WAS INITIATED BASED ON OTHER COMPLAINTS REPORTING THIS ISSUE.
MICS WAS NOT WORKING, HAD TO REPLACE BEFORE REAMING, LATER IN THE CASE INLINE OFFSET IMPACTOR STUCK IN EE AFTER IMPACTION. THA PROCEDURE DELAYED A FEW MINUTES TO OPEN UP ANOTHER MICS; OTHERWISE, NO FURTHER DELAY FOR THE STUCK IMPACTOR. SURGICAL DELAY WAS ABOUT 5 MINUTES. PER THE INVESTIGATION, A PIN WAS MISSING FROM THE HIP EE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 503053 | HIP END EFFECTOR, VARIABLE ANGLE | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | LOT 19060214 SN 1900368 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |