FDA Adverse Event
Malfunction
Summary report: N
INTRATEMP
MDR report key: 7650768
·
Received June 29, 2018
Report
- Report Number
- 7650768
- Event Type
- Malfunction
- Date Received
- June 29, 2018
- Date of Event
- May 31, 2018
- Report Date
- June 5, 2018
- Manufacturer
- MICROTEK MEDICAL, INC.
- Product Code
- LHC
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- WI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
THE BASIN FLUID WARMER SIDE POCKETS BECAME VERY HOT YET THE DIGITAL DISPLAY INDICATED 107 DEGREES F. THE TWO SIDE "BOTTLE POCKETS" ARE NOT TO EXCEED 110 DEGREES F. WE HAVE FOUND THAT THE BOTTLE TEMPERATURE CAN REACH 130 DEGREES F WHEN THE BOTTLE IS NOT RESTING OVER THE TEMPERATURE SENSOR. VARIOUS SIZE SOLUTION BOTTLES MAY BE PLACED INTO THESE POCKETS. THE SMALLER BOTTLES CAN BE PLACED IN THE POCKET SUCH THAT IT IS NOT RESTING OVER THE TEMPERATURE SENSOR. WHEN THE SENSOR IS NOT TOUCHING THE SIDE OF THE SOLUTION BOTTLE IT OVERHEATS THE BOTTLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 494676 | INTRATEMP | WARMER, IRRIGATION SOLUTION | LHC | MICROTEK MEDICAL, INC. | ITW6L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4015 DA |