FDA Adverse Event Malfunction Summary report: N

INTRATEMP

MDR report key: 7650768 · Received June 29, 2018

Report

Report Number
7650768
Event Type
Malfunction
Date Received
June 29, 2018
Date of Event
May 31, 2018
Report Date
June 5, 2018
Manufacturer
MICROTEK MEDICAL, INC.
Product Code
LHC
Product Problem
Yes
Report Source
User Facility report
Reporter Location
WI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

THE BASIN FLUID WARMER SIDE POCKETS BECAME VERY HOT YET THE DIGITAL DISPLAY INDICATED 107 DEGREES F. THE TWO SIDE "BOTTLE POCKETS" ARE NOT TO EXCEED 110 DEGREES F. WE HAVE FOUND THAT THE BOTTLE TEMPERATURE CAN REACH 130 DEGREES F WHEN THE BOTTLE IS NOT RESTING OVER THE TEMPERATURE SENSOR. VARIOUS SIZE SOLUTION BOTTLES MAY BE PLACED INTO THESE POCKETS. THE SMALLER BOTTLES CAN BE PLACED IN THE POCKET SUCH THAT IT IS NOT RESTING OVER THE TEMPERATURE SENSOR. WHEN THE SENSOR IS NOT TOUCHING THE SIDE OF THE SOLUTION BOTTLE IT OVERHEATS THE BOTTLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
494676 INTRATEMP WARMER, IRRIGATION SOLUTION LHC MICROTEK MEDICAL, INC. ITW6L

Patients

Seq Age Sex Outcome Treatment
1 4015 DA