TOSOH HLC-723G8 ANALYZER G8
Report
- Report Number
- 8031673-2018-00598
- Event Type
- Malfunction
- Date Received
- June 29, 2018
- Date of Event
- June 8, 2018
- Report Date
- September 18, 2019
- Manufacturer
- TOSOH CORPORATION
- Product Code
- LCP
- PMA / PMN Number
- K071132
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS CONDUCTED FOR SERIAL (B)(6), WHICH CONFIRMED THAT THERE WERE NO NONCONFORMANCES, FAILURES, DISCREPANCIES, OR MISSED STEPS DURING THE MANUFACTURING PROCESS THAT COULD BE RELATED TO THE REPORTED EVENT. TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE IMPORTER OR MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. DEVICE EVALUATION BY MANUFACTURER: A FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER TO ADDRESS THE REPORTED EVENT. DURING THE EVALUATION, FSE REPLACED THE UPTAKE CHECK VALVE AND THE STATOR FACE AND SEAL. FSE THEN PRIMED ALL FLUIDS AND RAN A DRAIN FLUSH. FSE RAN SAMPLES AND CONTROLS SEVERAL TIMES. THE COLUMN TEMPERATURE WAS 25.0. THE COLUMN COUNT AS ARRIVED WAS 4959. FSE REPLACED THE COLUMN THEN CALIBRATED AND RAN CONTROLS AND SAMPLES. CALIBRATION ( LOT ZS6002) WITH RESULTS OF 4.4, 4.4, 4.4, 9.0, 8.9. QUALITY CONTROL (QC) LEVEL 1 WAS 5.4 (RANGE 4.8 - 5.9); QC LEVEL 2 WAS 11.6 (RANGE 10.8 - 12.9). DATA INNOVATIONS (DI) LEVEL 1 WAS 5.4 AND DI LEVEL 2 WAS 11.5. PRECISION HAD A MEAN OF 5.670 WITH A STANDARD DEVIATION (SD) OF 0.048. THE COEFFICIENT OF VARIATION (%CV) OF 0.852. THE LIFE COUNT OF 09816. THE G8 INSTRUMENT WAS OPERATIONAL. NO FURTHER ACTION WAS REQUIRED BY THE FSE. A 13-MONTH COMPLAINT HISTORY REVIEW AND SERVICE HISTORY REVIEW FOR SIMILAR COMPLAINTS WAS PERFORMED FOR SERIAL NUMBER (B)(4) FROM (B)(6) 2017 THROUGH AWARE DATE (B)(6) 2018. THERE WERE NO SIMILAR COMPLAINTS IDENTIFIED DURING THE SEARCHED PERIOD. THE G8 VARIANT ANALYSIS MODE OPERATOR'S MANUAL UNDER CHAPTER 1, INTRODUCTION AND APPLICATIONS, STATES THE FOLLOWING: 1.8 LIMITATIONS OF THE PROCEDURE TOTAL AREA DILUTION STUDIES DEMONSTRATE THAT THE ASSAY IS LINEAR FROM A TOTAL AREA OF 500 TO 4000. HOWEVER, THE OPTIMUM TOTAL AREA IS 700 TO 3000. INTERPRETATION OF RESULTS WILL NOT BE REPORTED IF THE TOTAL AREA (TA) IS <500 WHICH CAN BE SEEN IN SEVERE ANEMIA. RESULTS WILL NOT BE REPORTED IF THE TA IS >4000 WHICH CAN BE SEEN IN POLYCYTHEMIA. (SEE "ABNORMAL RED CELL SURVIVAL IN PREVIOUS SECTION). THE OPTIMAL GOAL FOR TOTAL AREA IS BETWEEN 700-3000. HOWEVER A TA IN THE RANGE OF 500-4000 IS ACCEPTABLE AND REPORTABLE FOR WHOLE BLOOD SPECIMENS. CHAPTER 6 TROUBLESHOOTING STATES: 200 AREA LOW ERROR THREE SUCCESSIVE RESULTS BELOW THE LOWER LIMIT OF THE TOTAL AREA (50) OCCUR. IF THE ERROR MESSAGE IS PRESENT WHEN SUFFICIENT VOLUME OF SAMPLE IS SET IN THE RACK, THE PROBLEM MAY BE CAUSED BY AN EMPTY REAGENT (HEMOLYSIS & WASH SOLUTION). CHECK THE REMAINING VOLUME OF HEMOLYSIS & WASH SOLUTION AND START THE ASSAY AGAIN. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT WAS DUE TO FAILURE OR FAULT OF THE UPTAKE VALVE.
A CUSTOMER REPORTED GETTING ERROR 200 AREA LOW ERROR ON QUALITY CONTROL (QC) LEVEL 2 WITH THE G8 INSTRUMENT. THEY CUSTOMER MADE SEVERAL HEMOLYSATES AND A FRESH QC BOTTLE BUT ALL HAVE A LOW AREA. QC LEVEL 1 WAS ACCEPTABLE. TO TROUBLESHOOT, THE CUSTOMER RAN A WHOLE BLOOD DILUTION CUP FIVE TIMES AND 4 OUT OF 5 READ LOW TOTAL AREA ERROR. ADDITIONALLY, THE CUSTOMER RECEIVED A LOW TOTAL AREA ON QC DILUTION. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT, WHICH RESULTED IN DELAY IN REPORTING OF PATIENT RESULTS FOR HEMOGLOBIN A1C (HBA1C). THERE IS NO INDICATION OF ANY PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING OF PATIENT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 494612 | TOSOH HLC-723G8 ANALYZER G8 | G8 | LCP | TOSOH CORPORATION | G8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |