FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES

MDR report key: 7648686 · Received June 28, 2018

Report

Report Number
1917413-2018-02838
Event Type
Malfunction
Date Received
June 28, 2018
Date of Event
June 8, 2018
Report Date
September 13, 2018
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903679213
PMA / PMN Number
K945952
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: DUE TO AN IT ISSUE BEGINNING ON 7/3/2018, PREVIOUSLY FILED EMDRS DID NOT CONTAIN REQUIRED FIELDS. THIS SUPPLEMENTAL EMDR IS FILED TO PROVIDE THE FOLLOWING OMITTED FIELDS: DEVICE RETURNED TO MANUFACTURE: YES.

Additional Manufacturer Narrative · 0

BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. DRAW VOLUME TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND ALL SAMPLES MET THE REQUIRED SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. BASED ON EVALUATION OF THE CUSTOMER SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. A CAPA HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES WERE UNDERFILLED. NO REPORT OF SERIOUS INJURY OR MEDICAL INTERVENTION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES WERE UNDERFILLED. NO REPORT OF SERIOUS INJURY OR MEDICAL INTERVENTION.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 7256999, MEDICAL DEVICE EXPIRATION DATE: 12/31/2018, DEVICE MANUFACTURE DATE: 09/13/2017. MEDICAL DEVICE LOT #: 7180567, MEDICAL DEVICE EXPIRATION DATE: 10/31/2018, DEVICE MANUFACTURE DATE: 06/29/2017. MULTIPLE PMA / 510(K) #S. OTHER LISTED NUMBER IS K901449. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES WERE UNDERFILLED. NO REPORT OF SERIOUS INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
488546 BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. SEE H.10. 50382903679213

Patients

Seq Age Sex Outcome Treatment
1 Other