ARCHITECT I2000SR ANALYZER
Report
- Report Number
- 1628664-2018-00260
- Event Type
- Malfunction
- Date Received
- June 28, 2018
- Date of Event
- June 9, 2018
- Report Date
- July 30, 2018
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- PMA / PMN Number
- K983212
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
AN ABBOTT TECHNICAL SUPPORT SERVICE (TSS) INDIVIDUAL WAS DISPATCHED TO THE ACCOUNT AND FOUND A DAMAGED BOARD [BD, 36V MODULE ARTESYN/ASTEC PS (ROHS) PART 7-92695-02] ON THE ANALYZER. THE TSS NOTED SEVERAL MELTED COMPONENTS ON THIS BOARD. THE SMOKE WAS LIMITED TO THIS PART AND DID NOT SPREAD TO OTHER PARTS OF THE EQUIPMENT. A REPLACEMENT BOARD WAS ORDERED AND WAS INSTALLED BY THE CUSTOMER. THE ANALYZER WAS RETURNED TO NORMAL OPERATION. EVALUATION OF THE CUSTOMER ISSUE INCLUDED A COMPLAINT TEXT REVIEW, A SEARCH FOR SIMILAR COMPLAINTS, DEVICE HISTORY REVIEW FOR THE BOARD, LABELING REVIEW, AND AN INSTRUMENT SERVICE REVIEW. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER ISSUE. AN 85 MONTH SEARCH RETURNED NO ADDITIONAL COMPLAINTS FOR SMOKE EVENTS INVOLVING THIS BOARD [BD, 36V MODULE ARTESYN/ASTEC PS (ROHS) PART 7-92695-02]. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY ISSUES THAT MAY HAVE CAUSED THE CUSTOMER ISSUE. PRODUCT LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE, AS IT CONTAINS INSTRUCTIONS REGARDING ELECTRICAL SPECIFICATIONS AND REQUIREMENTS, ELECTRICAL HAZARDS, AND OF ELECTRICAL SAFETY FOR THE ARCHITECT SYSTEMS. ANALYZER SERIAL NUMBER SERVICE HISTORY WAS REVIEWED AND NO CONTRIBUTING FACTORS WERE FOUND. NO SUBSEQUENT REPORTS OF SMOKE/BURN ISSUES WERE IDENTIFIED. THE AFFECTED PART WAS NOT RETURNED. THE ISSUE WAS RESOLVED THROUGH PART REPLACEMENT. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS EVALUATION, NO SYSTEMIC ISSUE WAS IDENTIFIED AND NO PRODUCT DEFICIENCY WAS FOUND.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER AND ABBOTT TECHNICAL SUPPORT SERVICE (TSS) OBSERVED SMOKE ON THE 36 VOLT BOARD [BD, 36V MODULE ARTESYN/ASTEC PS (ROHS) PART 7-92695-02] OF THE ARCHITECT I2000SR ANALYZER. PARTS OF THE BOARD WERE MELTED. NO FIRE OR INJURY WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 488908 | ARCHITECT I2000SR ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | JJE | ABBOTT MANUFACTURING INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |