FDA Adverse Event Malfunction Summary report: N

PRIMUSRS

MDR report key: 7642962 · Received June 27, 2018

Report

Report Number
1119903-2018-00002
Event Type
Malfunction
Date Received
June 27, 2018
Date of Event
May 28, 2018
Report Date
May 29, 2018
Manufacturer
BTE TECHNOLOGIES, INC.
Product Code
ISD
PMA / PMN Number
K933611
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO INJURY RELATED TO THE SHAFT ISSUE HAS BEEN REPORTED. DESPITE THE DEVICE MALFUNCTIONING, BASED ON THE DESIGN OF THE PRIMUS AND THE NATURE OF THE FRACTURE, EVIDENCE SHOWS THAT A PATIENT OR USER WILL BE AT MINIMAL RISK FOR INJURY. BTE PERFORMED AN ANALYSIS REVIEWING THE SHAFT FAILURE CONDITIONS. INSPECTION OF CROSS-SECTIONS OF THE LOAD CELL SHAFTS REVEALED THAT THE SHAFT WILL WITHSTAND THE RATED LOADS UNLESS A UNIQUE AND RARE SET OF MANUFACTURING CONDITIONS ARE ENCOUNTERED. THE POSSIBILITY OF THESE CONDITIONS OCCURRING ONLY EXISTS ON A SMALL SUBSET OF SHAFTS. THE PROBABILITY THAT A SHAFT WILL FRACTURE OR SEPARATE IS REMOTE. IT HAS BEEN DETERMINED THAT NO REMEDIAL ACTION IS NECESSARY. REF. H3: THE LOAD CELL SHAFT FROM THE REPORTER FACILITY'S DEVICE HAS BEEN EVALUATED BASED ON EVIDENCE PROVIDED BY THE USER. THERE WAS NO NEED TO INSPECT THE ENTIRE PRIMUS SYSTEM ON THE USER SITE. SHAFTS FROM INVENTORY AND SHAFTS FROM OTHER DISTRIBUTED SYSTEMS HAVE BEEN ALSO EVALUATED. SUSPECT SHAFTS UNDERWENT LOAD TESTING AT BTE, AND ALL PARTS PASSED THE TEST.

Additional Manufacturer Narrative · 1

BTE PERFORMED AN ANALYSIS REVIEWING THE SHAFT FAILURE CONDITIONS. INSPECTION OF CROSS-SECTIONS OF THE LOAD CELL SHAFTS UNCOVERED THAT THE SHAFT WILL NOT BE ABLE TO WITHSTAND THE RATED LOADS ONLY WHEN A UNIQUE AND RARE SET OF MANUFACTURING CONDITIONS ARE ENCOUNTERED. THE POSSIBILITY OF THESE CONDITIONS OCCURRING ONLY EXISTS ON A SMALL SUBSET OF SHAFTS. THE PROBABILITY THAT A SHAFT WILL FRACTURE OR SEPARATE IS REMOTE.

Description of Event or Problem · 1

PRIMUS LOAD CELL SHAFT FRACTURED AND SEPARATED WHILE PATIENT WAS EXERCISING ON THE PRIMUS. NO INJURY OCCURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
485023 PRIMUSRS PRIMUS ISD BTE TECHNOLOGIES, INC. PRIMUSRS

Patients

Seq Age Sex Outcome Treatment
1