FDA Adverse Event Malfunction Summary report: N

90° CONTRA ANGLE DRIVER

MDR report key: 7638365 · Received June 26, 2018

Report

Report Number
0001032347-2018-00400
Event Type
Malfunction
Date Received
June 26, 2018
Date of Event
May 29, 2018
Report Date
November 28, 2018
Manufacturer
BIOMET MICROFIXATION
Product Code
HRS
PMA / PMN Number
PK142823
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. THE 90 DEGREE CONTRA ANGLE SCREWDRIVER (PART# 24-1189, LOT# 191750) AND CONTRA ANGLE TEMPORARY FIXATION SCREW (PART# 76-0017, LOT# 632390) WERE RETURNED IN A RED BIOHAZARD BAG. VISUAL INSPECTION SHOWED THAT THE PARTS SHOWED SIGNS OF WEAR, WITH SOME MINOR DISCOLORATION. THE INTERFACE BETWEEN THE SCREWDRIVER AND THE SCREW ALSO APPEARED RUSTY. THE TEMPORARY FIXATION SCREW COULD NOT BE RELEASED FROM THE SCREWDRIVER COLLET. THE SCREWDRIVER WAS NOT DISASSEMBLED DUE TO THE FACT THAT IT WAS RETURNED IN A BIOHAZARDOUS CONDITION. FURTHER FUNCTIONAL TESTS AND INSPECTIONS COULD NOT BE PERFORMED DUE TO THE BIO HAZARDOUS STATE OF THE PRODUCT. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. THERE ARE NO INDICATIONS OF MANUFACTURING DEFECTS. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: 0001032347-2018-00401-1.

Description of Event or Problem · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT (B)(4). MEDICAL PRODUCT: BIOMET MICROFIXATION RIBFIX BLU SYSTEM TEMPORARY FIXATION SCREW, CONTRA ANGLE CATALOG #: 76-0017, LOT #: 632390. THERAPY DATE: (B)(6) 2018. PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT; PLEASE SEE ASSOCIATED REPORT: 0001032347-2018-00401.

Description of Event or Problem · 1

IT WAS REPORTED DURING A RIB PLATING CASE A SHORT TEMPORARY PIN BECAME IRREVERSIBLY LODGED IN THE RIGHT ANGLE MANUAL DRIVER. ANOTHER SET WAS OPENED AND THE SURGEON USED THE CONTRA ANGLE DRIVER AND TEMPORARY FIXATION SCREW FROM THAT SET TO COMPLETE THE PROCEDURE. THE INCIDENT DID NOT CAUSE A SIGNIFICANT DELAY OR HARM THE PATIENT. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
480246 90° CONTRA ANGLE DRIVER SCREWDRIVER HRS BIOMET MICROFIXATION N/A 191750

Patients

Seq Age Sex Outcome Treatment
1