GIA
Report
- Report Number
- 2647580-2018-03149
- Event Type
- Malfunction
- Date Received
- June 26, 2018
- Date of Event
- June 8, 2018
- Report Date
- October 4, 2018
- Manufacturer
- US SURGICAL PUERTO RICO
- Product Code
- GDW
- UDI-DI
- 10884523007220
- PMA / PMN Number
- K111825
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
ADDITIONAL: EVALUATION SUMMARY: POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF ONE DEVICE WITH SINGLE USE LOADING UNIT. VISUAL INSPECTION NOTED THE SINGLE USE LOADING UNIT (SULU) WAS RECEIVED PARTIALLY APPLIED WITH THE INTERLOCK ENGAGED. MICROSCOPIC INSPECTION OF THE KNIFE BLADE DISPLAYED NO DAMAGE. THE SINGLE USE LOADING UNIT (SULU) WAS LOADED INTO A TEST DEVICE. THE SINGLE USE LOADING UNIT (SULU) AND INSTRUMENT WERE APPLIED TO TEST MEDIA WITH ACCEPTABLE RESULTS. RECORDS FROM EACH MANUFACTURING LOT ARE THOROUGHLY REVIEWED TO ENSURE THAT PRODUCTS ARE RELEASED MEETING ALL QUALITY RELEASE SPECIFICATIONS AT THE TIME OF MANUFACTURE. REPLICATION OF THE REPORTED CONDITION MAY OCCUR IF THE FIRING STROKE IS NOT COMPLETE AND THE FIRING KNOB IS NOT FULLY RETRACTED AFTER FIRING OR IF THE INSTRUMENT IS APPLIED AGAINST AN EXCESSIVE AMOUNT OF TISSUE DURING THE CLINICAL APPLICATION. IF THE FIRING KNOB IS NOT FULLY RETRACTED, DIFFICULTY IN UNLOADING THE LOADING UNIT MAY BE EXPERIENCED, AND MAY PREVENT FURTHER LOADING FROM OCCURRING. THE INSTRUCTIONS FOR USE (IFU) STATES THAT TO REMOVE THE SINGLE USE LOADING UNIT (SULU), SEPARATE THE INSTRUMENT HALVES. HOLDING THE CARTRIDGE-HALF OF THE INSTRUMENT, GRASP THE PROXIMAL END OF THE SINGLE USE LOADING UNIT (SULU) TABS AND PULL UP AND OUT TO REMOVE THE SINGLE USE LOADING UNIT (SULU) FROM THE INSTRUMENT. TO PLACE A NEW SINGLE USE LOADING UNIT (SULU) INTO THE INSTRUMENT, HOLD THE SINGLE USE LOADING UNIT (SULU) BY THE PROXIMAL END TABS AND INSERT INTO THE CARTRIDGE FORK AT A 30 TO 45 DEGREE ANGLE FROM THE DISTAL END DOWN UNTIL THE UNIT SNAPS INTO PLACE. REMOVE SHIPPING WEDGE AFTER SINGLE USE LOADING UNIT (SULU) IS FULLY LOADED. SINGLE USE LOADING UNIT (SULU) WILL ¿FLOAT¿ UP AND DOWN IN FINAL LOADED POSITION. THE ROOT CAUSE OF THE OBSERVED DAMAGE WAS MISUSE OF THE PRODUCT WHICH WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
ACCORDING TO THE REPORTER, IN ESOPHAGECTOMY PROCEDURE, HE WAS ABOUT TO FIRE THE DEVICE, BUT FOUND THAT THE KNIFE BLADE ON THE RELOAD WOULD NOT BE PUSHED OUT. THE SURGEON THEN USED ANOTHER RELOAD AND FOUND THE KNIFE BLADE WAS SMOOTH TO PUSH NOW AND WAS ABLE TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 482155 | GIA | STAPLE, IMPLANTABLE | GDW | US SURGICAL PUERTO RICO | GIA6038L | P7C1110X | 10884523007220 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |