FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PLUS PLASTIC WHOLE BLOOD TUBE

MDR report key: 7636245 · Received June 25, 2018

Report

Report Number
1917413-2018-03053
Event Type
Malfunction
Date Received
June 25, 2018
Date of Event
June 5, 2018
Report Date
September 6, 2018
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903678445
PMA / PMN Number
K981013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND UNDERFILL WAS NOT OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. BASED ON EVALUATION OF THE CUSTOMER PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS CONDUCTED AND UNDERFILL WAS NOT OBSERVED. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.

Additional Manufacturer Narrative · 0

DUE TO AN IT ISSUE BEGINNING ON 7/3/2018, PREVIOUSLY FILED EMDRS DID NOT CONTAIN REQUIRED FIELDS. THIS SUPPLEMENTAL EMDR IS FILED TO PROVIDE THE FOLLOWING OMITTED FIELDS: DEVICE RETURNED TO MANUFACTURE: YES.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND UNDERFILL WAS NOT OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. BASED ON EVALUATION OF THE CUSTOMER PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS CONDUCTED AND UNDERFILL WAS NOT OBSERVED. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BEFORE USE A BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES WAS FOUND WITH A "REDUCE IN VOLUME WAS NOTICED IN THE TUBES." THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION NEEDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BEFORE USE A BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES WAS FOUND WITH A "REDUCE IN VOLUME WAS NOTICED IN THE TUBES." THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION NEEDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BEFORE USE A BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES WAS FOUND WITH A "REDUCE IN VOLUME WAS NOTICED IN THE TUBES." THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION NEEDED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 7033587, EXPIRATION DATE: 2018-06-30, MANUFACTURE DATE: 2017-02-02. MEDICAL DEVICE LOT #: 7180947, EXPIRATION DATE: 2018-11-30, MANUFACTURE DATE:2017-06-29. MEDICAL DEVICE LOT #: 8011934, EXPIRATION DATE: 2019-05-31, MANUFACTURE DATE: 2018-01-11.

Description of Event or Problem · 1

IT WAS REPORTED THAT BEFORE USE A BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES WAS FOUND WITH A "REDUCE IN VOLUME WAS NOTICED IN THE TUBES." THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
475905 BD VACUTAINER® PLUS PLASTIC WHOLE BLOOD TUBE BLOOD COLLECTION TUBE JKA BECTON, DICKINSON & CO. SEE H.10 50382903678445

Patients

Seq Age Sex Outcome Treatment
1 Other