TITAN OTR NAR SCRO 18CM
Report
- Report Number
- 2125050-2018-00461
- Event Type
- Injury
- Date Received
- June 22, 2018
- Date of Event
- May 17, 2018
- Report Date
- July 13, 2018
- Manufacturer
- COLOPLAST A/S
- Product Code
- FHW
- PMA / PMN Number
- P000006
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
QUALITY WAS UNSUCCESSFUL IN SECURING THE DEVICE FOR EVALUATION. WITHOUT THE BENEFIT OF ANALYZING THE DEVICE, QUALITY CANNOT CONFIRM ANY OBSERVATIONS AND CANNOT COMMENT ON THE CONDITION OF THE PROSTHESIS. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. REVIEW OF NONCONFORMING REPORTS REVEALED NO NONCONFORMANCE WITH THIS LOT THAT WOULD HAVE CONTRIBUTED TO THE EVENT. A REVIEW OF THE COMPLAINT DATABASE REVEALED NO SIGNIFICANT TRENDS IN COMPLAINTS OF THIS TYPE FOR LOT 5851615. THE REVIEW OF THE DEVICE LOT HISTORY RECORDS BY QUALITY ASSURANCE INDICATES THIS LOT PASSED STERILITY TESTING PRIOR TO BEING RELEASED. BASED ON THIS KNOWLEDGE, QUALITY CONCLUDES THAT THE DEVICE WAS STERILE PRIOR TO THE DEVICE PACKAGING BEING OPENED AND THAT THE INFECTION ORIGINATED FROM SOURCE(S) OTHER THAN THE DEVICE. MANAGEMENT ROUTINELY REVIEWS EVENTS SUCH AS THIS AND MONITORS COMPLAINT LEVELS. ADDITIONALLY, EVENTS OF THIS TYPE ARE CAPTURED IN THE PRODUCT RISK DOCUMENTATION. BASED ON THIS INFORMATION, NO FURTHER CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. ONCE OUR EVALUATION IS COMPLETE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
ACCORDING TO THE AVAILABLE INFORMATION, INFECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471484 | TITAN OTR NAR SCRO 18CM | INFLATABLE PENILE PROSTHESIS | FHW | COLOPLAST A/S | QNR9181022 | 5851615 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Other |