DRIED POS MIC PANEL TYPE 20A
Report
- Report Number
- 2919016-2005-00032
- Event Type
- Malfunction
- Date Received
- November 9, 2005
- Date of Event
- October 13, 2005
- Report Date
- October 13, 2005
- Manufacturer
- DADE BEHRING INC.
- Product Code
- LTT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
REQUESTED CLINICAL ISOLATE FROM USER FOR EVALUATION. ROUTINE MONITORING OF COMPLAINT HISTORY AND PERFORMANCE TRENDS TO ENSURE PERFORMANCE IS WITHIN CLAIMS. EVAL CODES: CONCLUSIONS; OVERALL OXACILLIN PERFORMANCE OF DRIED POS PANELS IS ACCEPTABLE.
SINGLE ACCOUNT REPORTED TO DADE BEHRING THAT THEY OBSERVED A CLINICAL S. AUREOUS ISOLATE OXACILLIN MIC DISCREPANCY. THE ACCOUNT OBTAINED OXACILLIN-SUSCEPTIBLE RESULTS ON THE DRIED POS MIC PANEL TYPE 20A PANEL AND OXACILLIN-RESISTANT RESULTS ON A SECONDARY METHOD MRSA SCREEN AGAR) FOR THE CLINICAL ISOLATE. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES TO THE PT AS A RESULT OF THE FALSE SUSCEPTIBLE RESULTS BEING OBTAINED. CLINICAL ISOLATE NO LONGER AVAILABLE TO BE SUBMITTED BY THE ACCOUNT FOR TESTING BY DADE BEHRING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DRIED POS MIC PANEL TYPE 20A | ANTIMICROBIC SUSCEPT. TEST PANELS: GRAM POS. | LTT | DADE BEHRING INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |