DRIED POS MIC PANEL TYPE 20A

MDR report key: 761927 · Received November 9, 2005

Report

Report Number
2919016-2005-00032
Event Type
Malfunction
Date Received
November 9, 2005
Date of Event
October 13, 2005
Report Date
October 13, 2005
Manufacturer
DADE BEHRING INC.
Product Code
LTT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

REQUESTED CLINICAL ISOLATE FROM USER FOR EVALUATION. ROUTINE MONITORING OF COMPLAINT HISTORY AND PERFORMANCE TRENDS TO ENSURE PERFORMANCE IS WITHIN CLAIMS. EVAL CODES: CONCLUSIONS; OVERALL OXACILLIN PERFORMANCE OF DRIED POS PANELS IS ACCEPTABLE.

Description of Event or Problem · 1

SINGLE ACCOUNT REPORTED TO DADE BEHRING THAT THEY OBSERVED A CLINICAL S. AUREOUS ISOLATE OXACILLIN MIC DISCREPANCY. THE ACCOUNT OBTAINED OXACILLIN-SUSCEPTIBLE RESULTS ON THE DRIED POS MIC PANEL TYPE 20A PANEL AND OXACILLIN-RESISTANT RESULTS ON A SECONDARY METHOD MRSA SCREEN AGAR) FOR THE CLINICAL ISOLATE. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES TO THE PT AS A RESULT OF THE FALSE SUSCEPTIBLE RESULTS BEING OBTAINED. CLINICAL ISOLATE NO LONGER AVAILABLE TO BE SUBMITTED BY THE ACCOUNT FOR TESTING BY DADE BEHRING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DRIED POS MIC PANEL TYPE 20A ANTIMICROBIC SUSCEPT. TEST PANELS: GRAM POS. LTT DADE BEHRING INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 * Other