FDA Adverse Event Other Summary report: N

AXSYM CMV-G REAGENT KIT

MDR report key: 761898 · Received August 2, 2006

Report

Report Number
1415939-2006-00018
Event Type
Other
Date Received
August 2, 2006
Date of Event
July 19, 2006
Report Date
July 19, 2006
Manufacturer
ABBOTT LABORATORIES
Product Code
LSO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER STATES THAT ONE PT SAMPLE GENERATED AN INITIAL AXSYM CMV IGG ASSAY RESULT OF 73.5 AU/ML. THE SAMPLE WAS LATER RETESTED WITH LOT 31293M200 AND GENERATED AN AXSYM CMV IGG ASSAY RESULT OF 0 AU/ML. THE CUSTOMER THEN TESTED THE SAMPLE WITH A THIRD (DIFFERENT) LOT OF REAGENT AND GENERATED AN AXSYM CMV IGG ASSAY RESULT CLOSE TO THE FIRST RESULT OF 73.5 AU/ML (THE CUSTOMER DID NOT GIVE ANY INFO CONCERNING THE OTHER LOTS USED IN TESTING). THE CUSTOMER STATED THAT OVERALL, LOT 31293M200 GENERATES LOWER TEST RESULTS THAN OTHER LOTS USED IN THE LAB. CONTROLS HAVE REMAINED WITHIN SPECIFICATINS WITH ALL LOTS OF REAGENT IN USE. THERE IS NO IMPACT TO PT MGMT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AXSYM CMV-G REAGENT KIT IN VITRO TEST KIT FOR CMV ANTIBODIES LSO ABBOTT LABORATORIES NA 31293M200

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN AXSYM ANALYZER LIST#7A83-01