Narratives
ONE USED SAMPLE WAS RETURNED FOR EVALUATION. THE SUCTION CATHETER WAS NOT RETURNED WITH THE FLEX CONNECTOR SAMPLE. THE SAMPLE WAS IN GENERALLY CLEAN CONDITION, AND EXHIBITED JAGGED CRACKS. THE EDGES WERE EXAMINED UNDER MAGNIFICATION, AND HAD A SOMEWHAT SERRATED APPEARANCE. THE COMPLAINT IS CONFIRMED AS REPORTED. ALL INFORMATION REASONABLY KNOWN AS OF 15 NOV 2018 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC.. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE HALYARD HEALTH COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4).
CORRECTION: PER ADDITIONAL INFORMATION RECEIVED ON 21 JUN 2018 THE PRODUCT INFORMATION HAS BEEN UPDATED. ALL INFORMATION REASONABLY KNOWN AS OF 29 JUN 2018 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY HALYARD HEALTH REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO HALYARD HEALTH. HALYARD HEALTH HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE HALYARD HEALTH COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4).
THE PRODUCT INVOLVED IN THE REPORT HAS BEEN RETURNED AND IS BEING PROCESSED FOR EVALUATION. ALL INFORMATION REASONABLY KNOWN AS OF 24 SEP 2018 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY HALYARD HEALTH REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO HALYARD HEALTH. HALYARD HEALTH HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE HALYARD HEALTH COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT A HALYARD HEALTH PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.
THE ACTUAL COMPLAINT PRODUCT WAS NOT RETURNED FOR EVALUATION. ROOT CAUSE COULD NOT BE DETERMINED. THE DEVICE HISTORY RECORD FOR LOT M7263L602 WAS REVIEWED AND THE PRODUCT WAS PRODUCED ACCORDING TO PRODUCT SPECIFICATIONS. ALL INFORMATION REASONABLY KNOWN AS OF 19 JUN 2018 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY HALYARD HEALTH REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO HALYARD HEALTH. HALYARD HEALTH HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE HALYARD HEALTH COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). DEVICE NOT RETURNED.
IT WAS REPORTED THAT THERE WAS A CRACK IN THE FLEX CONNECTOR. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 457447 | CLOSED SUCTION CATHETER, DOUBLE SWIVEL ELBOW | VAP CLOSED SUCTION CATHETERS & ACCESSORIES | BSY | HALYARD HEALTH | 22108-5 | M7263L602 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |