BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE
Report
- Report Number
- 1917413-2018-00640
- Event Type
- Malfunction
- Date Received
- June 19, 2018
- Date of Event
- March 23, 2018
- Report Date
- July 30, 2018
- Manufacturer
- BECTON, DICKINSON & CO.
- Product Code
- JKA
- UDI-DI
- 50382903679626
- PMA / PMN Number
- K945952
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION SUMMARY: BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE CUSTOMER SAMPLES WERE EVALUATED ALONG WITH RETENTION SAMPLES OF THE INCIDENT LOT SELECTED FROM BD INVENTORY. THE CUSTOMER AND RETENTION SAMPLES WERE TESTED AND NO ISSUES RELATING TO ERRONEOUS RESULTS WERE OBSERVED AS ALL RESULTS DEMONSTRATED SATISFACTORY PERFORMANCE. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER AND RETAIN SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR ERRONEOUS RESULTS WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL SAMPLES MET THE REQUIRED SPECIFICATIONS. ROOT CAUSE DESCRIPTION: BASED ON THE INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED. THE PRODUCT WAS FOUND TO BE IN CONFORMANCE AND MEET RELEASE SPECIFICATIONS. RATIONALE: COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE HAD FALSELY ELEVATED TROPONIN LEVELS. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION.
FOUR POTENTIAL LOT NUMBERS WERE PROVIDED FOR THIS INCIDENT. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8032770, MEDICAL DEVICE EXPIRATION DATE: 2019-02-28, DEVICE MANUFACTURE DATE: 2018-02-01, MEDICAL DEVICE LOT #: 8011591, MEDICAL DEVICE EXPIRATION DATE: 2019-01-31, DEVICE MANUFACTURE DATE: 2018-01-11, MEDICAL DEVICE LOT #: 8011592, MEDICAL DEVICE EXPIRATION DATE: 2019-01-31, DEVICE MANUFACTURE DATE: 2018-01-11, MEDICAL DEVICE LOT #: 7340955, MEDICAL DEVICE EXPIRATION DATE: 2018-12-31, DEVICE MANUFACTURE DATE: 2017-12-06. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN (LH) BLOOD COLLECTION TUBE HAD FALSELY ELEVATED TROPONIN LEVELS. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 457018 | BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE | BLOOD COLLECTION TUBE | JKA | BECTON, DICKINSON & CO. | SEE H.10. | 50382903679626 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |