FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE

MDR report key: 7615206 · Received June 19, 2018

Report

Report Number
1917413-2018-00640
Event Type
Malfunction
Date Received
June 19, 2018
Date of Event
March 23, 2018
Report Date
July 30, 2018
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903679626
PMA / PMN Number
K945952
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE CUSTOMER SAMPLES WERE EVALUATED ALONG WITH RETENTION SAMPLES OF THE INCIDENT LOT SELECTED FROM BD INVENTORY. THE CUSTOMER AND RETENTION SAMPLES WERE TESTED AND NO ISSUES RELATING TO ERRONEOUS RESULTS WERE OBSERVED AS ALL RESULTS DEMONSTRATED SATISFACTORY PERFORMANCE. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER AND RETAIN SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR ERRONEOUS RESULTS WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL SAMPLES MET THE REQUIRED SPECIFICATIONS. ROOT CAUSE DESCRIPTION: BASED ON THE INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED. THE PRODUCT WAS FOUND TO BE IN CONFORMANCE AND MEET RELEASE SPECIFICATIONS. RATIONALE: COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE HAD FALSELY ELEVATED TROPONIN LEVELS. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION.

Additional Manufacturer Narrative · 1

FOUR POTENTIAL LOT NUMBERS WERE PROVIDED FOR THIS INCIDENT. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS:  MEDICAL DEVICE LOT #: 8032770, MEDICAL DEVICE EXPIRATION DATE: 2019-02-28, DEVICE MANUFACTURE DATE: 2018-02-01, MEDICAL DEVICE LOT #: 8011591, MEDICAL DEVICE EXPIRATION DATE: 2019-01-31, DEVICE MANUFACTURE DATE: 2018-01-11, MEDICAL DEVICE LOT #: 8011592, MEDICAL DEVICE EXPIRATION DATE: 2019-01-31, DEVICE MANUFACTURE DATE: 2018-01-11, MEDICAL DEVICE LOT #: 7340955, MEDICAL DEVICE EXPIRATION DATE: 2018-12-31, DEVICE MANUFACTURE DATE: 2017-12-06. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN (LH) BLOOD COLLECTION TUBE HAD FALSELY ELEVATED TROPONIN LEVELS. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457018 BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE BLOOD COLLECTION TUBE JKA BECTON, DICKINSON & CO. SEE H.10. 50382903679626

Patients

Seq Age Sex Outcome Treatment
1 Other