GALILEO ECHO
Report
- Report Number
- 1034569-2018-00143
- Event Type
- Malfunction
- Date Received
- June 18, 2018
- Date of Event
- April 23, 2018
- Report Date
- May 23, 2018
- Manufacturer
- IMMUCOR, INC.
- Product Code
- KSZ
- UDI-DI
- 10888234001584
- PMA / PMN Number
- BK070016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
IMMUCOR FIELD SERVICE WAS CALLED TO ASSESS THE GALILEO ECHO INSTRUMENT. FIELD SERVICE INSPECTED THE INSTRUMENT ON (B)(6) 2018 AND PERFORMED UNEXPECTED REACTION CHECKLIST. ALL TESTS PASSED EXCEPT FOR THE PERI-PUMP WHICH HAD A READING 9.29G. THE TECHNICIAN REPLACED THE TUBING AND THE TEST PASSED AT 10.9G. THE TECHNICIAN THEN RAN QC AND 7 PATIENT SAMPLES WHICH WERE ALL SUCCESSFUL. CUSTOMER WAS ABLE TO REPEAT SCREEN TEST OF SAMPLE ID (B)(4) ON (B)(6) 2018WITH CAPTURE-R READY SCREEN LOT NUMBER R968 AND CAPTURE-R INDICATOR CELLS LOT NUMBER 221085, OBTAINING 3+ RESULTS ON CELL 2 AS EXPECTED AFTER IMMUCOR FIELD SERVICE HAD CHANGED PUMP TUBING.
ON (B)(6) 2018, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE ANTIBODY SCREEN AND IDENTIFICATION, WHEN TESTED ON A GALILEO ECHO INSTRUMENT ON (B)(6) 2018.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 453217 | GALILEO ECHO | AUTOMATED BLOOD BANK SYSTEM | KSZ | IMMUCOR, INC. | 10888234001584 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR |