VITROS 5600 INTEGRATED SYSTEM
Report
- Report Number
- 1319681-2018-00055
- Event Type
- Malfunction
- Date Received
- June 18, 2018
- Date of Event
- March 15, 2018
- Report Date
- June 18, 2018
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT LOWER THAN EXPECTED VITROS CRBM RESULTS WERE OBTAINED DURING AN INTERNAL SPECIMEN MATRIX STUDY. THE TESTING WAS PERFORMED USING THREE DIFFERENT LOTS OF VITROS CRBM ON A VITROS 5600 INTEGRATED SYSTEM. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE CONSISTENT WITH AN INSTRUMENT ISSUE. DURING THE TIME INTERVAL THAT THE LOW CRBM RESULTS WERE OBTAINED, THE VITROS INSTRUMENT WAS INTERMITTENTLY POSTING IMMUNORATE WASH CONDITION CODES WHICH INDICATE THE SLIDE WAS NOT WASHED CORRECTLY. THE OPERATOR PERFORMED MAINTENANCE ON THE IMMUNO RATE SUBSYSTEM WHICH INCLUDED REPLACING THE TIP AND FORCING AIR THROUGH THE TUBING. THE MAINTENANCE ACTIONS RESOLVED THE IMMUNO RATE WASH CODES AND THE CRBM RESULTS OBTAINED POST MAINTENANCE WERE CONSISTENT WITH THE EXPECTED CRBM RESULTS.
AN ORTHO INTERNAL EMPLOYEE OBTAINED LOWER THAN EXPECTED VITROS CRBM RESULTS ON THREE SAMPLES FROM THE SAME WHOLE BLOOD POOL WHEN THE SAMPLES WERE PROCESSED USING THREE (3) SEPARATE LOTS OF VITROS CHEMISTRY PRODUCTS CRBM SLIDES ON A VITROS 5600 INTEGRATED SYSTEM. SAMPLE S-38 = 5.3 UG/ML VERSUS EXPECTED 13.6 UG/ML; SAMPLE NH-40 = 8.3, 9.2 UG/ML VERSUS EXPECTED 15.9 UG/ML; SAMPLE S-29 = 7.1 UG/ML VERSUS EXPECTED 10.2 UG/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. NO RESULTS WERE REPORTED TO A PHYSICIAN AS THE TESTING WAS INTERNAL TO ORTHO, HOWEVER; THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS COULD NOT BE AFFECTED OR WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED IN A HEALTH CARE SETTING. THERE WAS NO REPORTED ALLEGATION OF PATIENT HARM. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBER (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 454251 | VITROS 5600 INTEGRATED SYSTEM | CHEMISTRY ANALYZER | JJE | ORTHO-CLINICAL DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |