COLLEAGUE VOLUMETRIC INFUSION PUP
Report
- Report Number
- 6000001-2005-05340
- Event Type
- Malfunction
- Date Received
- October 19, 2005
- Date of Event
- September 1, 2005
- Report Date
- September 22, 2005
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-2/25/05-004-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE PUMP IS NOT AVAILABLE FOR EVALUATION. THE DEVICE HAS BEEN REQUESTED, BUT HAS NOT BEEN RECEIVED. SHOULD THE PUMP BE RECEIVED FOR EVALUATION, A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ADDITIONAL INFO BECOMES AVAILABLE. THE FAILURE CODE 570 INDICATES THAT THE BATTERY DISCHARGED TO A POTENTIALLY DAMAGING LEVEL. CAPA MDQ-CAPA-132 WAS OPENED AND IS INVESTIGATING REPORTS OF THE FAILURE CODE 570.
CUSTOMER REPORTED A FAILURE CODE 570:20. CUSTOMER REPORTED THERE WERE NO PT INJURIES ASSOCIATED WITH THIS REPORT AND THERE HAVE BEEN NO INCIDENT REPORTS FILED SINCE THE LAST BAXTER SERVICE EVENT. CUSTOMER DID NOT HAVE INFO WHETHER INCIDENT OCCURRED DURING PT INFUSION. ALTHOUGH BAXTER REQUESTED, NO ADDITIONAL INFO WAS PROVIDED REGARDING PT DEMOGRAPHICS, MEDICATION INVOLVED, OR DEVICE PROGRAMMING INFO. NO ADDITIONAL CONTACT INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | COLLEAGUE | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |