AMISTEM H HA COATED STD STEM SIZE 2
Report
- Report Number
- 3005180920-2018-00433
- Event Type
- Injury
- Date Received
- June 15, 2018
- Date of Event
- June 15, 2018
- Report Date
- June 15, 2018
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030804083
- PMA / PMN Number
- K093944
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER ON 07 JUNE 2018. HIP REVISION SURGERY IN A (B)(6) WOMAN. NO INFORMATION ON DATE OF PRIMARY SURGERY IS AVAILABLE. RADIOGRAPHIC IMAGE PROVIDED SHOW THE PRESENCE OF RADIOLUCENT LINES IN GRUEN ZONES 1 AND 7 AND REDUCED BONE DENSITY IN THE GREATER TROCHANTER REGION. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT FOLLOWING CEMENTLESS TOTAL HIP ARTHROPLASTY AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS FAILURE CANNOT BE DETERMINED. PRELIMINARY INVESTIGATION PERFORMED BY R&D PRODUCT MANAGER ON 15 JUNE 2018. FROM IMAGES AND X-RAY PRELIMINARY INVESTIGATION, FEMORAL STEM SHOWS SIGNS OF EXTRACTION PROCESS ON THE TAPER AND PROXIMAL NECK AREA. HYDROXYAPATITE LAYER HAS BEEN COMPLETELY ABSORBED DURING IN-SITU TIME. BATCH REVIEW PERFORMED ON 15 JUNE 2018. LOT 132630: 70 ITEMS MANUFACTURED AND RELEASED ON 02 SEPTEMBER 2013. EXPIRATION DATE: 31 JULY 2018. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 69 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
WE WERE INFORMED ON (B)(6) 2018 THAT A REVISION SURGERY WILL BE DONE ON (B)(6) 2018 DUE TO STEM LOOSENING. ON (B)(6) 2018 WE WERE INFORMED THAT THE REVISION SURGERY WAS PERFORMED (RIGHT SIDE) WAS DONE ON (B)(6) 2018. PRIMARY SURGERY PERFORMED IN 2013 (EXACT DATE NOT AVAILABLE AT THE MOMENT).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449462 | AMISTEM H HA COATED STD STEM SIZE 2 | CEMENTLESS HIP STEM | LZO | MEDACTA INTERNATIONAL SA | 132630 | 07630030804083 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention |