FDA Adverse Event Malfunction Summary report: N

ALINITY C PROCESSING MODULE

MDR report key: 7603696 · Received June 14, 2018

Report

Report Number
3002809144-2018-00061
Event Type
Malfunction
Date Received
June 14, 2018
Date of Event
May 21, 2018
Report Date
July 14, 2018
Manufacturer
ABBOTT GERMANY
Product Code
JJE
UDI-DI
00380740137380
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EVALUATION OF THE CUSTOMER ISSUE INCLUDED A COMPLAINT TEXT REVIEW, A SEARCH FOR SIMILAR COMPLAINTS, A LABELING REVIEW, INSTRUMENT LOG REVIEW, AND AN INSTRUMENT SERVICE REVIEW. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. LOG REVIEW DID NOT IDENTIFY ANY ISSUES THAT CONTRIBUTED TO THE CUSTOMER ISSUE. SERVICE HISTORY REVIEW IDENTIFIED NO CONTRIBUTING FACTORS TO THE CUSTOMER ISSUE. RETURNS WERE NOT AVAILABLE. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS EVALUATION, NO SYSTEMIC ISSUE WAS IDENTIFIED AND NO PRODUCT DEFICIENCY WAS FOUND. DATE OF EVENT WAS UPDATED AS MAY 21, 2018.

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. AN EVALUATION IS IN PROCESS.

Description of Event or Problem · 1

THE CUSTOMER OBSERVED MULTIPLE FALSELY DEPRESSED CALCIUM (CA) RESULTS AND A FALSELY DEPRESSED SODIUM (NA) RESULTS WHILE USING THE ALINITY C ANALYZER. THE FOLLOWING DATA WAS PROVIDED. CALCIUM (CA): (B)(6) INITIAL 3 MG/DL, REPEAT 4.3, 8.3, 9.2, 10.65 MG/DL; (B)(6) INITIAL 2.5 MG/DL, REPEAT 5, 7, 8, 9.32 MG/DL; (B)(6) INITIAL 3.7 MG/DL, REPEAT 5.7, 8.8 MG/DL. SODIUM (NA): (B)(6) INITIAL 145 MMOL/L, REPEAT 163, 163 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
446359 ALINITY C PROCESSING MODULE AUTOMATED CHEMISTRY ANALYZER JJE ABBOTT GERMANY 00380740137380

Patients

Seq Age Sex Outcome Treatment
1 CLINICAL CHEMISTRY CALCIUM| CLINICAL CHEMISTRY CALCIUM| LIST 07P57-30 LOT UNKNOWN| LIST 07P57-30 LOT UNKNOWN| SODIUM REAGENT LIST AND LOT UNKNOWN| SODIUM REAGENT LIST AND LOT UNKNOWN