TOSOH HLC-723G8 ANALYZER G8
Report
- Report Number
- 8031673-2018-00551
- Event Type
- Malfunction
- Date Received
- June 14, 2018
- Date of Event
- May 29, 2018
- Report Date
- June 14, 2018
- Manufacturer
- TOSOH CORPORATION
- Product Code
- LCP
- PMA / PMN Number
- K071132
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. DEVICE EVALUATION BY MANUFACTURER: A FIELD SERVICE ENGINEER (FSE) VISITED THE ACCOUNT TO ADDRESS THE REPORTED EVENT. THE FSE INSPECTED THE INSTRUMENT AND FOUND A LEAK AT THE SAMPLE LOOP. THE SAMPLE LOOP CONNECTOR WAS LOOSE AND FSE TIGHTENED THE CONNECTION. FSE VERIFIED PRECISION AND QUALITY CONTROLS WERE WITHIN RANGE. THE G8 INSTRUMENT WAS THEN OPERATIONAL IN ACCORDANCE WITH MANUFACTURER'S SPECIFICATIONS. NO FURTHER ACTION WAS REQUIRED BY THE FSE. A 13-MONTH COMPLAINT HISTORY REVIEW AND SERVICE HISTORY REVIEW FOR SIMILAR COMPLAINTS WAS PERFORMED FOR SERIAL NUMBER (B)(4) FROM 29-APR-2017 THROUGH AWARE DATE 29-MAY-2018. THERE WERE NO SIMILAR COMPLAINT(S) IDENTIFIED DURING THE SEARCHED PERIOD. THE G8 VARIANT ANALYSIS MODE OPERATOR'S MANUAL UNDER CHAPTER 6, TROUBLESHOOTING, STATES THE FOLLOWING: 6.3 ERROR MESSAGES: WHEN CONSULTING WITH TECHNICAL SUPPORT ABOUT A PROBLEM, PLEASE NOTE THE ERROR MESSAGE AND ERROR NUMBER. IN ADDITION, IF YOU FOLLOW THE SUGGESTED SOLUTIONS IN THIS SECTION AND ARE STILL UNABLE TO RESOLVE THE ERROR, OR IF YOU ENCOUNTER AN ERROR MESSAGE THAT IS NOT NOTED, CONTACT TECHNICAL SUPPORT. GENERAL ERROR MESSAGES: WITH THESE ERRORS, THE ASSAY STOPS AND THE ANALYZER IMMEDIATELY ENTERS STAND-BY STATE. THE 101 PRESSURE LOW: THE PRESSURE WILL NOT RISE BECAUSE THE PUMP IS UNABLE TO RUN DUE TO AIR BUBBLES IN THE PUMP CHECK VALVE. IF THE ELUTION BUFFER IS EMPTY, PLACE A NEW ELUTION BUFFER AND EXECUTE REAGENT CHANGE. NEXT, EXECUTE DRAIN FLUSH. SEE "CHAPTER 5 SECTION 5.5: PUMP AIR REMOVAL". EXECUTE MANUAL PUMPING USING THE PUMP KEY IN THE MAIN SCREEN (SECOND SCREEN), AND OPEN AND CLOSE THE DRAIN VALVE 2 OR 3 TIMES. IF THE PRESSURE RISES WHEN THE DRAIN VALVE IS CLOSED, THE OPERATION IS COMPLETE. IF THE PRESSURE STILL DOES NOT RISE OR STABILIZE, EXECUTE DRAIN FLUSH AGAIN. IN ADDITION, CONFIRM THAT THE DRAIN VALVE IS SECURELY CLOSED. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT WAS SAMPLE LOOP TUBING NOT TIGHT ON THE INJECTION VALVE.
A CUSTOMER REPORTED GETTING 101 PRESSURE LOW ERROR MESSAGE WITH THE G8 INSTRUMENT. THE CUSTOMER REPORTED THAT THEY TURNED ON THE INSTRUMENT TO PERFORM A DRAIN FLUSH AND NOTICED LEAKING FROM THE I-4 LINE (INJECTION VALVE). TECHNICAL SUPPORT (TS) INSTRUCTED THE CUSTOMER TO TIGHTEN THE CONNECTOR BUT THEY DID NOT HAVE A WRENCH TO DO SO. THE CUSTOMER REPORTED THAT THE TUBING WAS NOT INSERTED COMPLETELY IN THE CONNECTOR. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT, WHICH RESULTED IN DELAY IN REPORTING OF PATIENT RESULTS FOR HEMOGLOBIN A1C (HBA1C). THERE IS NO INDICATION OF ANY PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING OF PATIENT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 445761 | TOSOH HLC-723G8 ANALYZER G8 | G8 | LCP | TOSOH CORPORATION | G8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |