FDA Adverse Event
Injury
Summary report: N
HANA TABLE
MDR report key: 7600592
·
Received June 13, 2018
Report
- Report Number
- MW5077846
- Event Type
- Injury
- Date Received
- June 13, 2018
- Date of Event
- May 1, 2018
- Report Date
- June 4, 2018
- Manufacturer
- MIZUHO ORTHOPEDIC SYSTEMS INC.
- Product Code
- FQO
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- LA, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
PT WAS HAVING ANTERIOR HIP REPLACEMENT SURGERY AND THE NON OPERATIVE LEG FELL WHILE STILL IN THE LEG TABLE SUPPORT. THE DR DETERMINED THAT THE PT WAS NOT HARMED. MIZUHO OSI REPLACED THE PRESSURE ARM ASSEMBLY THAT WAS DAMAGED FROM THE LOCKING HANDLE EXCEEDING THE 2:00 POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 439956 | HANA TABLE | SURGICAL TABLE | FQO | MIZUHO ORTHOPEDIC SYSTEMS INC. | 6875 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention |