BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2018-01567
- Event Type
- Malfunction
- Date Received
- June 13, 2018
- Date of Event
- June 13, 2016
- Report Date
- May 30, 2018
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CI
- Reporter Occupation
- OTHER
Narratives
THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 5251002, MEDICAL DEVICE EXPIRATION DATE: 02/28/2017, DEVICE MANUFACTURE DATE: 09/08/2015. (B)(6). PMA / 510(K) #: THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US. BD HAD NOT RECEIVED SAMPLES, BUT PHOTOS WERE PROVIDED BY THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, RETENTION SAMPLES WERE SELECTED FOR EVALUATION, AND THE CUSTOMER'S INDICATED FAILURE MODE WAS NOT OBSERVED AS ALL SAMPLES MET SPECIFICATIONS. A REVIEW OF THE MANUFACTURING RECORDS WAS COMPLETED FOR THE INCIDENT LOT AND NO ISSUES WERE IDENTIFIED.
IT WAS REPORTED THAT BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES¿ PLASMA WAS MIXING WITH THE GEL AFTER CENTRIFUGATION AND PLUGGING THE NEEDLE. NO SERIOUS INJURY OR MEDICAL INTERVENTION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 437406 | BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | 5197045 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |