FDA Adverse Event
Malfunction
Summary report: N
R SERIES DEFIBRILLATOR
MDR report key: 7598204
·
Received June 13, 2018
Report
- Report Number
- 1220908-2018-01614
- Event Type
- Malfunction
- Date Received
- June 13, 2018
- Report Date
- May 24, 2018
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- DRT
- UDI-DI
- 00847946002695
- PMA / PMN Number
- K060559/P160
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MALFUNCTION WAS DUPLICATED AND THE ECG SHIELD ON THE ANALOG BOARD WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 439422 | R SERIES DEFIBRILLATOR | DEFIBRILLATOR/PACEMAKER | DRT | ZOLL MEDICAL CORPORATION | R SERIES | NA | 00847946002695 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |