FDA Adverse Event Malfunction Summary report: N

R SERIES DEFIBRILLATOR

MDR report key: 7598204 · Received June 13, 2018

Report

Report Number
1220908-2018-01614
Event Type
Malfunction
Date Received
June 13, 2018
Report Date
May 24, 2018
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
DRT
UDI-DI
00847946002695
PMA / PMN Number
K060559/P160
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MALFUNCTION WAS DUPLICATED AND THE ECG SHIELD ON THE ANALOG BOARD WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
439422 R SERIES DEFIBRILLATOR DEFIBRILLATOR/PACEMAKER DRT ZOLL MEDICAL CORPORATION R SERIES NA 00847946002695