FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM- BATTERY

MDR report key: 7597406 · Received June 13, 2018

Report

Report Number
3007042319-2018-02429
Event Type
Malfunction
Date Received
June 13, 2018
Date of Event
January 10, 2018
Report Date
April 14, 2020
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
UDI-DI
00888707000376
PMA / PMN Number
P100047
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. PRODUCT EVENT SUMMARY: THE BATTERY WAS RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICE IN RELATION TO THE REPORTED EVENT. A REVIEW OF THE MANUFACTURING DOCUMENTATION CONFIRMED THAT THE ASSOCIATED DEVICE MET ALL REQUIREMENTS FOR RELEASE. FAILURE ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE DEVICE PASSED EXTERNAL VISUAL EXAMINATION. FUNCTIONAL TESTING REVEALED A C ELL-UNDER-VOLTAGE FLAG AND PACK UNDER VOLTAGE FLAG, WHICH RENDERED THE BATTERY INOPERABLE. A CELL-UNDER-VOLTAGE FLAG IS TRIGGERED WHEN A CELL PAIR FALLS BELOW A VOLTAGE THRESHOLD. INTERNAL INSPECTION OF THE BATTERY REVEALED A LIFTED CELL WELD BETWEEN CELL PAIR 3 AND CELL PAIR 4, WHICH CONTRIBUTED TO THE CELL-UNDER-VOLTAGE FLAG AND PACK UNDER VOLTAGE. AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO A LIFTED CELL WELD. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

THIS REPORT CONTAINS AN UPDATE TO THE PRODUCT EVENT SUMMARY TO REFERENCE THE FORMAL INVESTIGATION ASSOCIATED WITH THE REPORTED FAILURE. THE PREVIOUSLY REPORTED FAILURE AND INVESTIGATION FINDINGS REMAIN UNCHANGED. MANUFACTURER CONTACT NAME AND DETAILS UPDATED. PRODUCT EVENT SUMMARY: BASED ON AN INVESTIGATION CONDUCTED UNDER SCAR2016039, THE ROOT CAUSE OF THE LIFTED CELL WELD CAN BE ATTRIBUTED TO TERMINAL WELDS EXPERIENCING ADDED STRESS AS A RESULT OF THE UNCONTROLLED BENDING OF THE TABS AND RESISTANCE WELDS OCCURRING ON WELD FREE ZONES. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE BATTERY PASSED EXTERNAL VISUAL EXAMINATION BUT FAILED FUNCTIONAL TESTING. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT¿S BATTERY WAS FLASHING RED AND WILL NOT CHARGE. BATTERY WAS EXCHANGED AND RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
440568 HEARTWARE VENTRICULAR ASSIST SYSTEM- BATTERY VENTRICULAR (ASSIST) BYPASS DSQ HEARTWARE, INC. 1650 00888707000376

Patients

Seq Age Sex Outcome Treatment
1