ALERE DETERMINE HIV 1/2 AG/AB COMBO
Report
- Report Number
- 1221359-2018-00315
- Event Type
- Malfunction
- Date Received
- June 13, 2018
- Date of Event
- August 21, 2017
- Report Date
- June 13, 2018
- Manufacturer
- ALERE SCARBOROUGH INC.
- Product Code
- MZF
- UDI-DI
- 07290015003735
- PMA / PMN Number
- BP120037-0
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
TESTING WAS PERFORMED ON RETAINED KIT LOT 170223 WITH THE FOLLOWING INTERNAL WHOLE BLOOD CONTROL SAMPLES: (B)(6), (B)(6), P24 (B)(4), AND (B)(6). THE KIT PERFORMED AS EXPECTED. ADDITIONALLY, QC DATA AND BATCH RECORDS FOR LOT 170223 WERE REVIEWED. RELEASE TESTING WAS VALID AND THE LOT WAS FOUND TO BE PERFORMING AS EXPECTED. A REVIEW OF THE COMPLAINTS REPORTED (B)(6) OR UNCONFIRMED (B)(6) RELATED TO LOT NUMBER 170223 SHOWED THAT THE COMPLAINT RATE IS (B)(4)%. THE EVIDENCE AVAILABLE DOES NOT INDICATE THAT THE PRODUCT IS PERFORMING OUTSIDE LABEL CLAIMS. IN HOUSE TESTING WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE RESULTS OBTAINED MAY POSSIBLY BE RELATED TO THE PATIENT SAMPLE. THE SAMPLE MAY HAVE CONTAINED SPECIFIC SUBSTANCES WHICH MAY HAVE AFFECTED THE RESULTS. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN LABELED CLAIMS.
CUSTOMER REPORTED A (B)(6) AB RESULT ON A BLOOD SAMPLE TESTED WITH THE ALERE DETERMINE HIV 1/2 AG/AB COMBO. ORAQUICK AND EIA 4TH GENERATION TESTS WERE (B)(6), RNA <20 COPIES/ML. PATIENT GENDER, PREGNANCY STATUS, TREATMENT AND PATIENT OUTCOMES WERE UNKNOWN. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY ALERE (B)(4); HOWEVER, THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN FDA OBSERVATION FROM AN FDA INSPECTION CONDUCTED FEBRUARY 26TH TO MARCH 2ND, 2018 AT ALERE (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 440408 | ALERE DETERMINE HIV 1/2 AG/AB COMBO | IVD FOR HIV AG/AB | MZF | ALERE SCARBOROUGH INC. | 170223 | 07290015003735 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |