ALERE DETERMINE HIV 1/2 AG/AB COMBO
Report
- Report Number
- 1221359-2018-00257
- Event Type
- Malfunction
- Date Received
- June 13, 2018
- Date of Event
- December 12, 2016
- Report Date
- June 13, 2018
- Manufacturer
- ALERE SCARBOROUGH INC.
- Product Code
- MZF
- UDI-DI
- 07290015003735
- PMA / PMN Number
- BP120037-0
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
AT THE TIME OF THE COMPLAINT AND INVESTIGATION, THE REPORTED LOT 160721 WAS EXPIRED AND UNABLE TO BE TESTED. HOWEVER, THE LOT WAS WITHIN EXPIRATION DATE WHEN USED FOR TESTING AT THE CUSTOMER SITE. THE MANUFACTURING BATCH RECORDS FOR LOT 160721 WERE REVIEWED. THIS LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED (B)(6) RELATED TO LOT NUMBER 160721 SHOWED THAT THE COMPLAINT RATE IS (B)(4). THE EVIDENCE AVAILABLE DOES NOT INDICATE THAT THE PRODUCT IS PERFORMING OUTSIDE LABEL CLAIMS. ALERE (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE RESULTS OBTAINED MAY POSSIBLY BE RELATED TO THE PATIENT SAMPLE. THE SAMPLE MAY HAVE CONTAINED SPECIFIC SUBSTANCES WHICH MAY HAVE AFFECTED THE RESULTS. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN LABELED CLAIMS.
A (B)(6) RESULT WAS REPORTED ON A BLOOD SAMPLE TESTED WITH THE ALERE DETERMINE HIV 1/2 AG/AB COMBO. THE PATIENT WAS CONFIRMED (B)(6). THERE IS INSUFFICIENT INFORMATION TO DETERMINE IF A MALFUNCTION OCCURRED. THE PATIENT GENDER, PREGNANCY STATUS, TREATMENT AND PATIENT OUTCOME WERE UNKNOWN. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY ALERE (B)(4), INC.; HOWEVER, THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED FEBRUARY 26TH TO MARCH 2ND, 2018 AT ALERE (B)(4), INC. ((B)(4)).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 439591 | ALERE DETERMINE HIV 1/2 AG/AB COMBO | IVD FOR HIV AG/AB | MZF | ALERE SCARBOROUGH INC. | 160721 | 07290015003735 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |