ALERE DETERMINE HIV 1/2 AG/AB COMBO
Report
- Report Number
- 1221359-2018-00289
- Event Type
- Malfunction
- Date Received
- June 13, 2018
- Date of Event
- August 3, 2017
- Report Date
- June 13, 2018
- Manufacturer
- ALERE SCARBOROUGH INC.
- Product Code
- MZF
- UDI-DI
- 07290015003735
- PMA / PMN Number
- BP120037-0
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
TESTING WAS PERFORMED AT ALERE (B)(4) ON RETAINED KIT LOT 160918 WITH THE FOLLOWING INTERNAL WHOLE BLOOD CONTROL SAMPLES: HIV-1 POSITIVE, HIV-2 POSITIVE, P24 POSITIVE, AND HIV NEGATIVE. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING BATCH RECORDS FOR LOT 160918 WERE REVIEWED. THIS LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED FALSE POSITIVE OR UNCONFIRMED FALSE POSITIVE RELATED TO LOT NUMBER 160918 SHOWED THAT THE COMPLAINT RATE IS (B)(4). THE EVIDENCE AVAILABLE DOES NOT INDICATE THAT THE PRODUCT IS PERFORMING OUTSIDE LABEL CLAIMS. ALERE SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE RESULTS OBTAINED MAY POSSIBLY BE RELATED TO THE PATIENT SAMPLE. THE SAMPLE MAY HAVE CONTAINED SPECIFIC SUBSTANCES WHICH MAY HAVE AFFECTED THE RESULTS. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN LABELED CLAIMS.
A (B)(6) RESULT WAS REPORTED WITH A VENOUS WHOLE BLOOD SAMPLE ON THE ALERE DETERMINE¿ HIV-1/2 AG/AB COMBO TEST. THE TEST WAS REPEATED AND REPORTED AS (B)(6). BASED ON THE INFORMATION AVAILABLE IN THE CASE, IT APPEARS A MALFUNCTION MAY HAVE OCCURRED WITH ONE OF THE TWO CONFLICTING ALERE DETERMINE HIV TESTS.THE PATIENT'S TREATMENT AND PATIENT OUTCOME WERE UNKNOWN. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY ALERE (B)(4).; HOWEVER, THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED FEBRUARY 26TH TO MARCH 2ND , 2018 AT ALERE (B)(4)
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 438251 | ALERE DETERMINE HIV 1/2 AG/AB COMBO | IVD FOR HIV AG/AB | MZF | ALERE SCARBOROUGH INC. | 160918 | 07290015003735 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |