FDA Adverse Event Malfunction Summary report: N

SEE H10.

MDR report key: 7593076 · Received June 12, 2018

Report

Report Number
2015691-2018-02251
Event Type
Malfunction
Date Received
June 12, 2018
Date of Event
May 22, 2018
Report Date
May 22, 2018
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THROUGH FURTHER INVESTIGATION, IT WAS LEARNED THAT THE SUBJECT DEVICE HAD CALCIFICATION ON THE LEAFLETS. THE PATIENT UNDERWENT SURGICAL AORTIC VALVE REPLACEMENT AS THE PATIENT WAS NOT A CANDIDATE FOR TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR). THE AORTIC WAS TOO SMALL FOR A TAVR VALVE. CALCIFICATION PLAYS A MAJOR ROLE IN THE FAILURE OF BIOPROSTHETIC HEART VALVES. CALCIFICATION OF VALVES OCCURS AS A PROGRESSIVE, TIME-DEPENDENT PROCESS. TISSUE VALVE CALCIFICATION IS INITIATED PRIMARILY WITHIN RESIDUAL CELLS THAT HAVE BEEN DEVITALIZED. INITIAL CALCIFICATION DEPOSITS EVENTUALLY ENLARGE AND GROW INTO A MASS, WHICH STIFFEN AND WEAKEN THE TISSUE AND THEREBY CAUSE THE PROSTHESIS TO MALFUNCTION. THE MINERALIZATION OF A BIOMATERIAL IS GENERALLY ENHANCED AT THE SITES OF INTENSE MECHANICAL DEFORMATIONS GENERATED BY MOTION, SUCH AS THE POINTS OF FLEXION IN HEART VALVES. ULTIMATELY, THE RESULT OF CALCIFICATION IS VALVE FAILURE DUE TO TEARING OR STENOSIS. MANY FACTORS CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION. THESE INCLUDE PATIENT FACTORS (AGE, DISEASE STATE, PHARMACOLOGICAL INTERVENTION, ETC.), MECHANICAL STRESS RELATED TO THE VALVE'S HEMODYNAMIC PERFORMANCE, AND GLUTARALDEHYDE FIXATION OF TISSUE. OF THESE, THE FIXATION PROCESS IS A RELATIVELY MINOR CONTRIBUTOR TO CALCIFICATION FOR EDWARDS' TISSUE VALVES DUE TO ANTI-CALCIFICATION TREATMENTS DURING MANUFACTURING. THOUGH NUMEROUS STUDIES HAVE BEEN CONDUCTED ON PREVENTIVE CALCIFICATION STRATEGIES IN BIOPROSTHETIC HEART VALVES, THE CAUSES OF CALCIFICATION ARE NOT FULLY UNDERSTOOD AND THERE ARE STILL NO MECHANISMS OR MEDICAL THERAPIES WHICH FULLY PREVENT BIOPROSTHESES FROM CALCIFYING.

Additional Manufacturer Narrative · 1

THIS MODEL IS NOT SOLD OR MARKETED IN THE U.S. HOWEVER, IT IS SIMILAR TO DEVICE: MODEL 2800; BRAND NAME: CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; PMA NUMBER P860057/S001. ADDITIONAL MANUFACTURER NARRATIVE: THE SUBJECT DEVICE IS NOT AVAILABLE FOR EVALUATION, AS IT REMAINS IMPLANTED IN THE PATIENT. THE CLINICAL OBSERVATION WAS UNABLE TO BE CONFIRMED. THE DEVICE HISTORY RECORD (DHR) WAS NOT REVIEWED AS THE REPORTED EVENT DOES NOT ALLEGE A MALFUNCTION THAT COULD BE RELATED TO A MANUFACTURING DEFICIENCY AND/OR ONE WAS NOT CONFIRMED THROUGH INVESTIGATION. BIOPROSTHETIC TISSUE VALVES CAN DETERIORATE WITH TIME AND EVENTUALLY FAIL CONTRIBUTING TO REGURGITATION AND/OR STENOSIS. THERE CAN BE A NUMBER OF POTENTIAL KNOWN AND UNKNOWN PATIENT RELATED CONTRIBUTING FACTORS. STRUCTURAL VALVE DETERIORATION (SVD), A COMMON REASON FOR BIOPROSTHESIS EXPLANT OR REOPERATION, ENCOMPASSES MULTIPLE FAILURE MODES, INCLUDING CALCIFIC AND NON-CALCIFIC DEGENERATION, DEHISCENCE, CUSP THICKENING OR FIBROSIS, OR A COMBINATION OF THESE. SUCH FAILURE MODES, OCCURRING SINGULARLY OR CONCOMITANTLY, MAY CONTRIBUTE TO STENOSIS AND/OR REGURGITATION. ALTERNATIVELY, NONSTRUCTURAL DYSFUNCTION (NSVD) MAY ALSO PLAY A ROLE IN THE DEVELOPMENT OF VALVULAR STENOSIS. A MANUFACTURING NON-CONFORMANCE WAS NOT IDENTIFIED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED; HOWEVER, IT IS LIKELY THAT PATIENT RELATED FACTORS AND THE PROGRESSION OF THE UNDERLYING VALVULAR DISEASE PATHOLOGY CONTRIBUTED TO THE EVENT. THERE WAS NO INDICATION OR ALLEGATION OF A DEVICE MALFUNCTION CONTRIBUTING TO THE EVENT. THE INSTRUCTIONS FOR USE (IFU) HAS BEEN REVIEWED AND NO INADEQUACIES WERE IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL EVENTS THROUGH THE USE OF EDWARDS QUALITY SYSTEMS. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO FURTHER CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED AT THIS TIME.

Description of Event or Problem · 1

EDWARDS RECEIVED INFORMATION THAT A PATIENT WITH A 19MM BIOPROSTHETIC AORTIC VALVE WHICH WAS IMPLANTED 13 YEARS AND 8 MONTHS AGO, IS BEING CONSIDERED FOR SURGICAL INTERVENTION DUE TO AORTIC STENOSIS. THE PATIENT PRESENTED WITH SHORTNESS OF BREATH. SHE HAS A SMALL AORTA AND SMALL AORTIC ROOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435167 SEE H10. HEART-VALVE, REPLACEMENT DYE EDWARDS LIFESCIENCES 290019MM

Patients

Seq Age Sex Outcome Treatment
1 79 YR