5 ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 5 ML SYRINGE, 10 USP UNITS/ML
Report
- Report Number
- 2134319-2018-00059
- Event Type
- Injury
- Date Received
- June 11, 2018
- Date of Event
- February 19, 2018
- Report Date
- September 7, 2018
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FOZ
- UDI-DI
- 30382903065111
- PMA / PMN Number
- K011967
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
PRIOR TO THE REPORTED COMPLAINT OF INFECTION EVALUATED IN MPS-18-1248-SA, THERE WAS NO TREND FOR INFECTION CASES REPORTED FOR FRANKLIN PRODUCT. THIS COMPLAINT IS PART OF ANEW TREND WHICH BEGAN IN APRIL 2018. CAPA 350041 WAS INITIATED TO ADDRESS THIS ISSUE. ALTHOUGH A DEFINITIVE LOT # WAS NOT PROVIDED, A GENERAL DHR REVIEW WAS PERFORMED OF THE POTENTIAL LOTS LISTED. THERE WERE NO DEVIATIONS OR NON-CONFORMANCES THAT OCCURRED DURING THE MANUFACTURE OF THE POTENTIAL LOTS LISTED. THE STERILITY LOT TESTING RESULTS FOR THE POTENTIAL LOTS WERE ALL ACCEPTABLE (NO GROWTH EXHIBITED). NO SAMPLES PROVIDED. A REVIEW OF ALL LOT STERILITY TESTING PERFORMED FOR PRODUCT RELEASED BETWEEN APRIL 2015 & MAY 2018 CONFIRMED THAT NO ORGANISM GROWTH WAS IDENTIFIED FOR ANY LOTS RELEASED DURING THE SHELF-LIFE OF THIS PRODUCT. INVESTIGATION CONCLUSION: THE ROOT CAUSE ANALYSIS OF THE REPORTED INFECTION CASES UNDER CAPA 350041 HAS NOT IDENTIFIED A DIRECT CAUSATION BETWEEN THE INFECTIONS AND THE BD FRANKLIN PRODUCT.
IT WAS REPORTED THAT A 4 YEAR OLD FEMALE PATIENT WITH A DIAGNOSIS OF RESPIRATORY FAILURE AND SEPSIS AND A PICC LINE IN PLACE HAD A POSITIVE BLOOD CULTURE FOR SERRATIA MARCESCENS ON (B)(6) 2018. THE PATIENT WAS TREATED AND DISCHARGED. THE PFGE PATTERNS FOR THIS INCIDENT ARE INDISTINGUISHABLE FROM THE ORIGINAL CO OUTBREAK PATTERN. THE DEVICE IMPLICATED WITH THIS INCIDENT IS A 5 ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 5 ML SYRINGE, 10 USP UNITS/ML. *OUT OF AN ABUNDANCE OF CAUTION AND IN THE INTEREST OF PUBLIC HEALTH, BD VOLUNTARILY RECALLED CERTAIN LOTS OF BD POSIFLUSH¿ HEPARIN LOCK FLUSH AND BD¿ PRE-FILLED NORMAL SALINE FLUSH SYRINGES DUE TO A POTENTIAL FOR CONTAMINATION WITH SERRATIA MARCESCENS BACTERIUM. BD WAS NOTIFIED BY THE U.S. FOOD AND DRUG ADMINISTRATION (FDA) AND CENTERS FOR DISEASE CONTROL AND PREVENTION (CDC) ABOUT A POTENTIAL EPIDEMIOLOGICAL LINK BETWEEN CATHETER RELATED BLOOD STREAM INFECTIONS AND THE S. MARCESCENS BACTERIUM. SPECIFICALLY, THE FDA AND CDC IDENTIFIED A POTENTIAL CONNECTION BETWEEN REPORTS OF INFECTION IN A SMALL NUMBER OF PATIENTS CAUSED BY S. MARCESCENS ACROSS MULTIPLE STATES. CDC¿S INITIAL INVESTIGATION FOUND THAT AFFECTED PATIENTS HAD RECEIVED TREATMENT USING CERTAIN BD FLUSH PRODUCTS. TO DATE, THERE IS NO EVIDENCE OF BD FLUSH PRODUCT TESTING POSITIVE FOR THIS BACTERIUM. INVESTIGATIONS ARE ONGOING BY BD, FDA, AND CDC.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT A (B)(6) PATIENT WITH A DIAGNOSIS OF RESPIRATORY FAILURE AND SEPSIS AND A PICC LINE IN PLACE HAD A POSITIVE BLOOD CULTURE FOR SERRATIA MARCESCENS ON (B)(6) 2018. THE PATIENT WAS TREATED AND DISCHARGED. THE PFGE PATTERNS FOR THIS INCIDENT ARE INDISTINGUISHABLE FROM THE ORIGINAL CO OUTBREAK PATTERN. THE DEVICE IMPLICATED WITH THIS INCIDENT IS A 5 ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 5 ML SYRINGE, 10 USP UNITS/ML. OUT OF AN ABUNDANCE OF CAUTION AND IN THE INTEREST OF PUBLIC HEALTH, BD VOLUNTARILY RECALLED CERTAIN LOTS OF BD POSIFLUSH¿ HEPARIN LOCK FLUSH AND BD¿ PRE-FILLED NORMAL SALINE FLUSH SYRINGES DUE TO A POTENTIAL FOR CONTAMINATION WITH SERRATIA MARCESCENS BACTERIUM. BD WAS NOTIFIED BY THE U.S. FOOD AND DRUG ADMINISTRATION (FDA) AND CENTERS FOR DISEASE CONTROL AND PREVENTION (CDC) ABOUT A POTENTIAL EPIDEMIOLOGICAL LINK BETWEEN CATHETER RELATED BLOOD STREAM INFECTIONS AND THE S. MARCESCENS BACTERIUM. SPECIFICALLY, THE FDA AND CDC IDENTIFIED A POTENTIAL CONNECTION BETWEEN REPORTS OF INFECTION IN A SMALL NUMBER OF PATIENTS CAUSED BY S. MARCESCENS ACROSS MULTIPLE STATES. CDC¿S INITIAL INVESTIGATION FOUND THAT AFFECTED PATIENTS HAD RECEIVED TREATMENT USING CERTAIN BD FLUSH PRODUCTS. TO DATE, THERE IS NO EVIDENCE OF BD FLUSH PRODUCT TESTING POSITIVE FOR THIS BACTERIUM. INVESTIGATIONS ARE ONGOING BY BD, FDA, AND CDC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 427839 | 5 ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 5 ML SYRINGE, 10 USP UNITS/ML | PREFILLED HEPARIN FLUSH SYRINGE | FOZ | BECTON DICKINSON MEDICAL SYSTEMS | 726471N | 30382903065111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 YR | Required Intervention |