FDA Adverse Event Malfunction Summary report: N

ALERE DETERMINE HIV 1/2 AG/AB COMBO

MDR report key: 7589358 · Received June 11, 2018

Report

Report Number
1221359-2018-00203
Event Type
Malfunction
Date Received
June 11, 2018
Date of Event
May 22, 2017
Report Date
June 11, 2018
Manufacturer
ALERE SCARBOROUGH INC.
Product Code
MZF
UDI-DI
07290015003735
PMA / PMN Number
BP120037-0
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING WAS PERFORMED ON RETAINED KIT LOTS 161218 AND 170223 WITH THE FOLLOWING INTERNAL WHOLE BLOOD CONTROL SAMPLES: (B)(6). THE KITS PERFORMED AS EXPECTED. ADDITIONALLY, QC DATA AND BATCH RECORDS FOR LOT 161218 AND 170223 WERE REVIEWED. RELEASE TESTING WAS VALID AND THE LOTS WERE FOUND TO BE PERFORMING AS EXPECTED. A REVIEW OF THE COMPLAINTS REPORTED (B)(6) OR UNCONFIRMED (B)(6) RELATED TO LOT NUMBER 161218 SHOWED THAT THE COMPLAINT RATE IS (B)(4). A REVIEW OF THE COMPLAINTS REPORTED (B)(6) OR UNCONFIRMED (B)(6) RELATED TO LOT NUMBER 170223 SHOWED THAT THE COMPLAINT RATE IS (B)(4). THE EVIDENCE AVAILABLE DOES NOT INDICATE THAT THE PRODUCT IS PERFORMING OUTSIDE LABEL CLAIMS. IN HOUSE TESTING WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE RESULTS OBTAINED MAY POSSIBLY BE RELATED TO THE PATIENT SAMPLE. THE SAMPLE MAY HAVE CONTAINED SPECIFIC SUBSTANCES WHICH MAY HAVE AFFECTED THE RESULTS. THE AVAILABLE EVIDENCE SUGGESTS THAT THESE DEVICE LOTS ARE PERFORMING WITHIN LABELED CLAIMS.

Description of Event or Problem · 1

CUSTOMER REPORTED A (B)(6) AB RESULT ON A CAPILLARY WHOLE BLOOD SAMPLE TESTED WITH THE ALERE DETERMINE HIV 1/2 AG/AB COMBO. CUSTOMER REPEATED THE PATIENT'S TEST WITH A SECOND LOT WHICH ALSO GAVE A (B)(6) AB RESULT. QUEST (B)(6) AG AND ABS 4TH GENERATION TEST, WITH REFLEXES, WAS (B)(6). THERE IS INSUFFICIENT INFORMATION TO DETERMINE IF A MALFUNCTION OCCURRED. PATIENT GENDER, PREGNANCY STATUS, TREATMENT AND PATIENT OUTCOMES WERE UNKNOWN. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY ALERE (B)(4), INC.; HOWEVER, THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED FEBRUARY 26TH TO MARCH 2ND, 2018 AT ALERE (B)(4).