MENTOR SMOOTH ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2018-03590
- Event Type
- Injury
- Date Received
- June 11, 2018
- Date of Event
- May 18, 2018
- Report Date
- May 21, 2018
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001270
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
ON 6/27/2018, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. MENTOR ALSO BECAME AWARE THAT THE REPLACEMENT DEVICES WERE THE FOLLOWING: (LEFT) MENTOR SMOOTH ROUND MODERATE PROFILE 350CC SALINE CATALOG: 3501655, LOT: 7508732, SN: (B)(4) AND (RIGHT) MENTOR SMOOTH ROUND MODERATE PROFILE 350CC SALINE CATALOG: 3501655, LOT: 7385379, SN: (B)(4). DEVICE INVESTIGATION SUMMARY: UPON RECEIPT BY MENTOR, THE DEVICE CONTAINED NO FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE. NO FOREIGN MATERIAL WAS OBSERVED ON THE SHELL SURFACE. EVALUATION OF THE DEVICE REVEALED A RENT MEASURING APPROXIMATELY 0.4 CM WITHIN A CREASE ON THE POSTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT OF DEFLATION WAS CONFIRMED. MENTOR PERFORMS 100% INSPECTION AND TESTING OF ALL DEVICES PRIOR TO RELEASE, PRODUCT EVALUATION CONCLUDED THAT THE RENT AND CREASE OCCURRED SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CONCOMITANT PRODUCTS: MENTOR SMOOTH ROUND MODERATE PROFILE 375CC SALINE CATALOG: 3501660 LOT: 5915094 SN: (B)(4).
IT WAS REPORTED THAT A (B)(6) -YEAR OLD CAUCASIAN FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH MENTOR SMOOTH ROUND MODERATE PROFILE 375CC SALINE BREAST PROSTHESES. DEFLATION OF THE LEFT BREAST PROSTHESIS WAS DETERMINED. AS A RESULT, THE PATIENT UNDERWENT REMOVAL ON (B)(6) 2018.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 430441 | MENTOR SMOOTH ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 5915094 | 00081317001270 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR | Required Intervention |