FDA Adverse Event Malfunction Summary report: N

VOLISTA STANDOP

MDR report key: 7586464 · Received June 11, 2018

Report

Report Number
9710055-2018-00043
Event Type
Malfunction
Date Received
June 11, 2018
Date of Event
May 17, 2018
Report Date
June 26, 2018
Manufacturer
MAQUET SAS
Product Code
FSY
PMA / PMN Number
K130513
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET SAS, PARC DE LIMÈRE, AVENUE DE LA POMME DE PI ORLÉANS CEDEX 2, FRANCE 45074 EXEMPTION # E2018005. GETINGE USA SALES, LLC, 45 BARBOUR POND DRIVE, WAYNE, NJ 07470. CONTACT PERSON:(B)(4). THE ISSUE IS BEING INVESTIGATED BY MANUFACTURING SITE.

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET SAS, PARC DE LIMÈRE, AVENUE DE LA POMME DE PI ORLÉANS CEDEX 2, FRANCE 45074 EXEMPTION # E2018005 GETINGE USA SALES, LLC (B)(6). MAQUET SAS BECAME AWARE OF AN INCIDENT WITH A SURGICAL LIGHT VOLISTA STANDOP DEVICE. AS IT WAS STATED, CRACKS HAVE BEEN FOUND ON THE FORK. THE ISSUE OCCURRED BEFORE THE OPERATION HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION. IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE SURGICAL LIGHT DID NOT MEET ITS SPECIFICATION AND IT CONTRIBUTED TO EVENT. AT THE TIME WHEN THE EVENT OCCURRED THE DEVICE WAS NOT BEING USED FOR PATIENT TREATMENT. DURING THE INVESTIGATION IT WAS FOUND THAT THE REPORTED SCENARIO HAS NEVER LEAD TO SERIOUS INJURY OR WORSE, TO DEATH. THE CRACK APPEARS DUE TO THE DETACHMENT OF THE BONDED PARTS AND IS ONLY LOCATED ON THE OUTER COATING. THE CAUSE OF THE COATING DAMAGE CORRESPONDS TO AN EXCESSIVE GAP BETWEEN BOTH MECHANICAL PARTS REVEALED BEFORE THE COATING MANUFACTURING PROCESS. MAQUET SAS LAUNCHED THE CAPA 2015-03 AND THE DESIGN CHANGE REQUEST E160103 TO IMPROVE THE METHOD FOR ASSEMBLING THE BRACKETS REPLACING THE BONDING METHOD BY A WELDING PROCESS. THE BRACKET ASSEMBLED WITH THE WELDING PROCESS IS AVAILABLE SINCE THE BEGINNING OF 2017.

Description of Event or Problem · 0

ON (B)(6) 2018 MAQUET SAS BECAME AWARE OF AN INCIDENT WITH ONE OF SURGICAL LIGHTS- VOLISTA STANDOP. AS IT WAS STATED, CRACKS HAVE BEEN FOUND ON THE FORK. THE ISSUE OCCURRED BEFORE THE OPERATION HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION. (B)(4).

Description of Event or Problem · 0

MANUFACTURER REFERENCE NUMBER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
429489 VOLISTA STANDOP LIGHT, SURGICAL, CEILING MOUNTED FSY MAQUET SAS

Patients

Seq Age Sex Outcome Treatment
1