ALERE DETERMINE HIV 1/2 AG/AB COMBO
Report
- Report Number
- 1221359-2018-00153
- Event Type
- Malfunction
- Date Received
- June 8, 2018
- Date of Event
- March 16, 2017
- Report Date
- June 8, 2018
- Manufacturer
- ALERE SCARBOROUGH INC.
- Product Code
- MZF
- UDI-DI
- 07290015003735
- PMA / PMN Number
- BP120037-0
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
IN-HOUSE TESTING WAS PERFORMED ON RETAINED KIT LOT 161027 WITH THE FOLLOWING INTERNAL SERUM/PLASMA CONTROL SAMPLES: (B)(6). ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING BATCH RECORDS FOR LOT 161027 WERE REVIEWED. THIS LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED (B)(6) OR UNCONFIRMED (B)(6) RELATED TO LOT 161027 SHOWED THAT THE COMPLAINT RATE IS (B)(4). THE EVIDENCE AVAILABLE DOES NOT INDICATE THAT THE PRODUCT IS PERFORMING OUTSIDE LABEL CLAIMS. AN EXACT ROOT CAUSE OF THE REPORTED ISSUE WAS UNABLE TO BE DETERMINED. THE RESULT OBTAINED MAY POSSIBLY BE RELATED TO THE PATIENT SAMPLE. THE SAMPLE MAY HAVE CONTAINED SPECIFIC SUBSTANCES WHICH MAY HAVE AFFECTED THE RESULTS. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN LABEL CLAIMS.
A (B)(6) AG RESULT WAS REPORTED ON A SERUM SAMPLE TESTED WITH THE ALERE DETERMINE HIV 1/2 AG/AB COMBO. THE PATIENT WAS CONFIRMED (B)(6) BY (B)(6) NAAT. THERE IS INSUFFICIENT INFORMATION TO DETERMINE IF A MALFUNCTION OCCURRED. THE PATIENT GENDER, PREGNANCY STATUS, TREATMENT AND PATIENT OUTCOME WERE UNKNOWN. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY ALERE (B)(4), INC.; HOWEVER, THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED FEBRUARY 26TH TO MARCH 2ND, 2018 AT ALERE (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 424937 | ALERE DETERMINE HIV 1/2 AG/AB COMBO | IVD FOR HIV AG/AB | MZF | ALERE SCARBOROUGH INC. | 161027 | 07290015003735 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |