FDA Adverse Event Injury Summary report: N

CERVICAL PLATE VARIABLE SELF-TAPPING SCREW DIAM. 4X14MM (2X)

MDR report key: 7582505 · Received June 8, 2018

Report

Report Number
3005180920-2018-00409
Event Type
Injury
Date Received
June 8, 2018
Date of Event
May 16, 2018
Report Date
July 10, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
KWQ
UDI-DI
07630030840333
PMA / PMN Number
K140361
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR ON 11 JUNE 2018: REVISION SURGERY OCCURRED 3 YEARS AFTER PRIMARY PROCEDURE DUE TO PSEUDARTHROSIS. INCOMPLETE FUSION IS A POSSIBLE ADVERSE EVENTS AND MAY BE DUE TO DIFFERENT FACTORS EITHER MECHANICAL OR PATIENT RELATED. ON THE BASIS OF THE INFORMATION PROVIDED IT IS NOT POSSIBLE TO DETERMINE WHEN THE SCREW BREAKAGE OCCURRED AND ITS RELATION WITH PSEUDOARTHROSIS.

Additional Manufacturer Narrative · 1

THREE PIECES OF THE LOT 141356 AND A PIECE OF THE LOT 141188 REFERENCE 03.70.240 (K140361); CERVICAL PLATE L2X40MM WERE REMOVED DUE TO LEFT ARM AND NECK PAIN 3 YEARS AND 3 MONTHS AFTER PRIMARY. THE PATIENT HAS PSEUDOARTHROSIS. A PIECE OF THE LOT 141356 WAS FOUND BROKEN BY THE SURGEON. BATCH REVIEW PERFORMED ON 05 JUNE 2018; LOT 141356: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 22 OCTOBER 2014 EXPIRATION DATE: 2019-08-31 NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN COMPLAINING OF LEFT ARM AND NECK PAIN 3 YEARS AND 3 MONTHS AFTER PRIMARY. THE PATIENT HAS PSEUDOARTHROSIS. THE SURGEON REMOVED MEDACTA SCREWS AND 1 PLATE. ONE SCREW COULD HAVE NOT BEEN REMOVED AS THE HEAD WAS FOUND BROKEN DURING REVISION. THE SURGEON WAS ABLE TO RECOVER THE HEAD OF THE SCREW. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
425546 CERVICAL PLATE VARIABLE SELF-TAPPING SCREW DIAM. 4X14MM (2X) VARIABLE SELF-TAPPING SCREW KWQ MEDACTA INTERNATIONAL SA 141356 07630030840333

Patients

Seq Age Sex Outcome Treatment
1 76 YR Required Intervention