FDA Adverse Event Other Summary report: N

ARCHITECT I2000 ANALYZER

MDR report key: 758022 · Received October 19, 2005

Report

Report Number
1628664-2005-00042
Event Type
Other
Date Received
October 19, 2005
Date of Event
September 27, 2005
Report Date
October 17, 2005
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO RESULTS OR CONCLUSIONS AT THIS TIME, THE INVESTIGATION IS ON-GOING.

Description of Event or Problem · 1

THE ACCOUNT STATED A FALSE NEGATIVE ANTI-HCV RESULT WAS OBTAINED ON THE ARCHITECT I2000 SYSTEM. THE SAMPLE WAS FROM AN ALIQUOT AND HAD BEEN CENTRIFUGED AT HIGH SPEED AND RUN IN DUPLICATE WITH ONE REPLICATE GIVING A STRONG REACTIVE RESULT (9.28 S/CO) AND THE OTHER REPLICATE GIVING A NONREACTIVE RESULT (0.09 S/CO). BOTH SPECIMEN TUBES WERE RERUN AND BOTH GAVE REACTIVE ARCHITECT ANTI-HCV RESULTS (10.72, 10.67 S/CO). NO ARCHITECT I2000 INSTRUMENT ERRORS WERE NOTED DURING PROCESSING. THE NONREACTIVE ARCHITECT I2000 ANTI-HCV RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. NO PATIENT INFORMATION IS AVAILABLE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000 ANALYZER AUTOMATED IMMUNOASSAY ANALYZER JJE ABBOTT MANUFACTURING, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN ARCHITECT ANTI-HCV ASSAY, LIST 6C37