FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® RAPID SERUM TUBE (RST) BLOOD COLLECTION TUBES THROMBIN

MDR report key: 7576405 · Received June 6, 2018

Report

Report Number
2243072-2018-01280
Event Type
Malfunction
Date Received
June 6, 2018
Date of Event
February 24, 2018
Report Date
June 15, 2018
Manufacturer
BECTON DICKINSON
Product Code
JKA
UDI-DI
50382903687744
PMA / PMN Number
K101502
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

MEDICAL DEVICE TYPE UPDATED TO: JKA. 510(K)# UPDATED TO: K101502.

Additional Manufacturer Narrative · 1

IN THIS MDR, BD CORPORATE HEADQUARTERS IN (B)(4) HAS BEEN LISTED AS (B)(4) IS AN OEM MANUFACTURING SITE. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCE DURING MANUFACTURING OF THE PRODUCT. AS NO SAMPLES OR PHOTOS WERE RECEIVED FOR EVALUATION, THE CUSTOMER'S INDICATED FAILURE MODE WAS NOT OBSERVED BY BD. AS THERE WAS NO SAMPLE OR PHOTO AVAILABLE FOR EVALUATION, A ROOT CAUSE COULD NOT BE DETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOR THE BD VACUTAINER® RAPID SERUM TUBE (RST) BLOOD COLLECTION TUBES THROMBIN THERE ARE DISCREPANCIES WITH LD RESULTS. NO REPORT OF INJURY OR MEDICAL INTERVENTION. NO REPORT OF BLOOD EXPOSURE TO MUCOUS MEMBRANE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418456 BD VACUTAINER® RAPID SERUM TUBE (RST) BLOOD COLLECTION TUBES THROMBIN BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON DICKINSON 170306 50382903687744

Patients

Seq Age Sex Outcome Treatment
1 Other