FDA Adverse Event Malfunction Summary report: N

HANDPIECE MICS

MDR report key: 7572816 · Received June 6, 2018

Report

Report Number
3005985723-2018-00338
Event Type
Malfunction
Date Received
June 6, 2018
Date of Event
May 18, 2018
Report Date
September 25, 2018
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486030193
PMA / PMN Number
K170581
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

FOLLOW-UP #1 AND FINAL REPORT SUBMITTED. REPORTED EVENT: IT WAS REPORTED THAT THE MICS HANDLE FELL DURING THE CASE. PRODUCT EVALUATION AND RESULTS: THE PRODUCT WAS UNAVAILABLE FOR INSPECTION AS THE PRODUCT WAS NOT RETURNED. PRODUCT HISTORY REVIEW: REVIEW OF DEVICE HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED UNDER LOT K07GK AND (B)(4) INCLUDING 4200667 WERE ACCEPTED INTO FINAL STOCK ON 6/14/16. REVIEW OF QT16-06-0063 REVEALED THAT THE NON CONFORMANCE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLAINT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 209063, LOT NUMBER 42030416 SHOWS NO ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: THE FAILURE MODE WAS CONFIRMED BASED ON THE PICTURES PROVIDED THE PRODUCT WAS NOT AVAILABLE FOR INSPECTION. IF ADDITIONAL INFORMATION IS RECEIVED THE COMPLAINT WILL BE REOPENED. CORRECTIVE ACTION/PREVENTIVE ACTION: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN AN NC AND CAPA ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. THIS IS NC (B)(4) AND CAPA (B)(4). DEVICE NOT RETURNED.

Description of Event or Problem · 0

SURGEON PLACED MICS BACK IN HOLSTER POSITION. HANDLE AND SCREW FELL OUT. SURGICAL DELAY APPROXIMATELY 15 MINUTES. MICS HANDPIECE REPLACED. TKA PROCEDURE.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

SURGEON PLACED MICS BACK IN HOLSTER POSITION. HANDLE AND SCREW FELL OUT. SURGICAL DELAY APPROXIMATELY 15 MINUTES. MICS HANDPIECE REPLACED. TKA PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
414947 HANDPIECE MICS STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 420304164200667 00848486030193

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization